Uniform prostate imaging and spectroscopy at 7 T: comparison between a microstrip array and an endorectal coil

  • Utrecht University
    ,
  • Research not related to TU/e workplace (outside TU/e employment)
Output:
Contribution to journal
Article
Academic
Peer-review

Information

Output type

Output:
Contribution to journal
Article
Academic
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 358-365 (8 pages)

Journal (Volume, Issue Number)

NMR in Biomedicine (Volume 24, Issue 4)

Publication milestones

  • Published - 05/2011

Publication status

Published - 05/2011

ISSN

0952-3480

Abstract

An endorectal coil and an eight-element microstrip array were compared for prostate imaging at 7 T. An extensive radiofrequency safety assessment was performed with the use of finite difference time domain simulations to determine safe scan parameters. These simulations showed that the endorectal coil can deliver substantially more B(1)(+) to the prostate than can the microstrip array within the specific absorption rate safety guidelines. However, the B(1)(+) field of the endorectal coil is very inhomogeneous, which makes the use of adiabatic pulses compulsory for T(1) - or T(2) -weighted imaging. As a consequence, a full prostate examination is only possible in a feasible amount of time when the microstrip array is used for T(1) - and T(2) -weighted imaging, whereas the endorectal coil is required for spectroscopic imaging. The pulse parameters were optimised within the specific absorption rate guidelines and thereafter used to provide a good illustration of the possibilities of prostate imaging at 7 T.

Metrics

Metrics

Scopus
citations
36
SciVal
FWCI
1.59
SciVal
Author count
9
SciVal
Paper percentile
82
SciVal
citations
36

PlumX, opens in new tab

Captures
38
Citation count
36