The use of metal sublaminar wires in modern growth-guidance scoliosis surgery: A report of 4 cases and literature review
- R. Bogie(corresponding author),
- ,
- S. N. Koole,
- L. W. van Rhijn,
- P. C. Willems
- Sint Anna Hospital,
- Maastricht University Medical Center
Output:
Contribution to journal
Review article
Academic
Peer-reviewOpen access
Information
Output type
Output:
Contribution to journal
Review article
Academic
Peer-reviewOriginal language
EnglishPages from-to (Number of pages)
Pages 182-188 (7 pages)Journal (Volume, Issue Number)
International Journal of Spine Surgery (Volume 14, Issue 2)Publication milestones
- Published - 2020
Publication status
Published - 2020
ISSN
2211-4599Abstract
<p>Background: To avoid early fusion and allow residual growth of the spine in early onset scoliosis (EOS) treatment, growth-guided scoliosis surgery can be performed. Four patients with EOS are presented in which a growth-guidance instrumentation is used with sliding titanium (Ti) sublaminar cables. Residual growth of the spine can be preserved using metal sublaminar wiring; however, several drawbacks of this technique and type of material are illustrated. Methods: Four patients with progressive neuromuscular scoliosis were treated with a posterior stabilization. A fusionless growth-guidance instrumentation was used consisting of a combination of lumbar pedicle screws and sliding Ti sublaminar cables along cobalt chrome rods. Results: In 2 cases, the described growth-guidance technique provided sufficient stability and correction of the curvature with preservation of growth. In 2 patients, the instrumentation failed due to upper thoracic sublaminar wire breakage. The ongoing abrasion of the rod-wire interface caused severe metallosis. In these cases, a debridement and revision surgery was performed with partial fusion of the spine. Conclusions: Growth-guidance techniques with sliding metal sublaminar wires seem to be a valuable solution for the preservation of spinal growth in EOS surgery. High curvatures, however, have a higher chance of failure and demand for more corrective strength and support of the instrumentation. The use of metal sublaminar wires in a ''sliding'' instrumentation can lead to early breakage and metallosis. Level of Evidence: 4, case series. Clinical Relevance: Surgeons should be aware of possible complications associated with the use of metal laminar wires in spinal fusion and growth-guidance scoliosis surgery. The implementation of materials containing higher fatigue strength and lower friction properties (eg, UHMWPE wires) may avoid these potential complication risks.</p>
Bibliographical Note
Publisher Copyright:
© International Society for the Advancement of Spine Surgery
Documents and Links
Metrics
Metrics
SciVal
FWCI
0.28
SciVal
Author count
5
SciVal
Paper percentile
41
SciVal
citations
9
Scopus
citations
9
PlumX, opens in new tab
Citation count
9
Captures
14
