The role of stationary phases in reversed-phase liquid chromatography in the application of solvent optimization procedures
- M.A. Straten, van,
- R.J. Lynch,
- H.A. Claessens
Output:
Contribution to journal
Article
Academic
Open access
Information
Output type
Output:
Contribution to journal
Article
Academic
Original language
EnglishPages from-to (Number of pages)
Pages 143-150 (7 pages)Journal (Volume, Issue Number)
LC-GC International : the Magazine of Separation Science (Volume 8, Issue 3)Publication milestones
- Published - 1995
Publication status
Published - 1995
ISSN
0895-5441Abstract
The development of liquid chromatographic separation procedures, which meet high standards of reliability and ruggedness, is of increasing importance. A number of software programs are available to make method development in laboratory practice less time consuming and also less dependent on the degree of knowledge of the analyst. Method development, as well as the subsequent routine analysis, generally start from the assumption of a constant column quality (that is, selectivity). In this study the role of the column in method development in reversed-phase liquid chromatography was investigated. It is shown that the influence of nominally identical stationary phases from different manufacturers on the results of method development procedures is significant.
This also implies that the transferability of chromatographic separation conditions between "identical" columns is poor.
Documents and Links
Final published version, 1.72 MB
Metrics
Metrics
Download statistics
Download count
80