Quantification of bone microarchitecture and strength with photon-counting detector CT at different radiation doses: an ex-vivo and in-vivo comparison with HR-pQCT

  • ,
  • VieCuri Medical Center
    ,
  • Maastricht School for Nutrition and Translational Research in Metabolism
    ,
  • Erasmus University Medical Center
    ,
  • Maastricht University Medical Center
Output:
Contribution to journal
Article
Academic
Peer-review

Open access

Information

Output type

Output:
Contribution to journal
Article
Academic
Peer-review

Original language

English

Article number

112630

Journal (Volume, Issue Number)

European Journal of Radiology (Volume 196)

Publication milestones

  • E-pub ahead of print - 03/01/2026
  • Published - 03/2026

Publication status

Published - 03/2026

ISSN

0720-048X

Abstract

PURPOSE: To compare trabecular microarchitecture measurements at the distal radius and tibia from photon-counting detector CT (PCD-CT) at varying radiation doses with high-resolution peripheral quantitative CT (HR-pQCT).

METHODS: Two intact wrist and two intact ankle specimens from an 88-year-old man were scanned with PCD-CT at radiation doses of 2.5, 5, 10, and 20 mGy and with HR-pQCT. Additionally, clinical in-vivo HR-pQCT and PCD-CT scans at the radius and tibia were acquired of a 40-year-old woman with osteoporosis. After bone segmentation, the segmented PCD-CT and HR-pQCT scans were three-dimensionally registered. Cubic volumes (edge length: ex-vivo 6 mm, in-vivo 5 mm) were defined at corresponding locations in the PCD-CT and HR-pQCT scans based on the three-dimensional registration. For each cube, trabecular volume fraction (Tb.BV/TV), thickness (Tb.Th), number (Tb.N), separation (Tb.Sp), and heterogeneity (Tb.1/N.SD) were quantified and compared between corresponding PCD-CT and HR-pQCT cubes.

RESULTS: Ex-vivo, linear correlation coefficients (R 2) between PCD-CT and HR-pQCT were 0.85-0.97 at 2.5 mGy and remained stable with increasing radiation dose for all parameters except Tb.1/N.SD. For Tb.1/N.SD, R 2 increased between 2.5 and 5 mGy and remained stable at higher doses. At each radiation dose, Tb.BV/TV, Tb.N, and Tb.Th values were higher and Tb.Sp and Tb.1/N.SD lower on PCD-CT than on HR-pQCT. In-vivo, R 2 was 0.89-0.95 (radius) and 0.82-0.97 (tibia).

CONCLUSIONS: PCD-CT strongly correlated with HR-pQCT in trabecular microarchitecture measurements at the distal radius and tibia at low, clinically acceptable, radiation dose. Between-modality differences in microarchitecture values are likely related to chosen image analysis settings.