Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve: The RESOLVE study
- Allen Jeremias,
- Akiko Maehara,
- Philippe Généreux,
- Kaleab N. Asrress,
- Colin Berry,
- B. (Bernard) Bruyne, de
Output:
Contribution to journal
Article
Academic
Peer-reviewInformation
Output type
Output:
Contribution to journal
Article
Academic
Peer-reviewOriginal language
DutchPages from-to (Number of pages)
Pages 1253-1261 (9 pages)Journal (Volume, Issue Number)
Journal of the American College of Cardiology (Volume 63, Issue 13)Publication milestones
- Published - 2014
Publication status
Published - 2014
ISSN
0735-1097Abstract
Objectives This study sought to examine the diagnostic accuracy of the instantaneous wave-free ratio (iFR) and resting distal coronary artery pressure/aortic pressure (Pd/Pa) with respect to hyperemic fractional flow reserve (FFR) in a core laboratory-based multicenter collaborative study. Background FFR is an index of the severity of coronary stenosis that has been clinically validated in 3 prospective randomized trials. iFR and Pd/Pa are nonhyperemic pressure-derived indices of the severity of stenosis with discordant reports regarding their accuracy with respect to FFR. Methods iFR, resting Pd/Pa, and FFR were measured in 1,768 patients from 15 clinical sites. An independent physiology core laboratory performed blinded off-line analysis of all raw data. The primary objectives were to determine specific iFR and Pd/Pa thresholds with ≥90% accuracy in predicting ischemic versus nonischemic FFR (on the basis of an FFR cut point of 0.80) and the proportion of patients falling beyond those thresholds. Results Of 1,974 submitted lesions, 381 (19.3%) were excluded because of suboptimal acquisition, leaving 1,593 for final analysis. On receiver-operating characteristic analysis, the optimal iFR cut point for FFR ≤0.80 was 0.90 (C statistic: 0.81 [95% confidence interval: 0.79 to 0.83]; overall accuracy: 80.4%) and for PdPa was 0.92 (C statistic: 0.82 [95% confidence interval: 0.80 to 0.84]; overall accuracy: 81.5%), with no significant difference between these resting measures. iFR and PdPa had ≥90% accuracy to predict a positive or negative FFR in 64.9% (62.6% to 67.3%) and 48.3% (45.6% to 50.5%) of lesions, respectively. Conclusions This comprehensive core laboratory analysis comparing iFR and PdPa with FFR demonstrated an overall accuracy of ∼80% for both nonhyperemic indices, which can be improved to ≥90% in a subset of lesions. Clinical outcome studies are required to determine whether the use of iFR or PdPa might obviate the need for hyperemia in selected patients. © 2014 by the American College of Cardiology Foundation.
Metrics
Metrics
Scopus
citations
346
SciVal
FWCI
11.96
SciVal
Author count
25
SciVal
Paper percentile
99
SciVal
Top percentile
1
SciVal
citations
346
Download statistics
Download count
2
PlumX, opens in new tab
Citation count
351
Captures
230
Mentions
2