Modeling a heart pump
- V. Creigen,
- L. Ferracina,
- A.V. Hlod,
- S. Mourik, van,
- K. Sjauw,
- V. Rottschäfer
Output:
Chapter in Book/Report/Conference proceeding
Conference contribution
Academic
Peer-reviewInformation
Output type
Output:
Chapter in Book/Report/Conference proceeding
Conference contribution
Academic
Peer-reviewOriginal language
EnglishPages from-to (Number of pages)
Pages 7-25Publication milestones
- Published - 2007
Publication status
Published - 2007
Place of publication
UtrechtPublisher
Utrecht University, NetherlandsHost publication title
Proceedings 58th European Study Group Mathematics with Industry (ESGI58/SWI2007), 29 January - 2 February 2007, Utrecht, The NetherlandsHost publication editors
- R.H. Bisseling
- K. Dajani
- T.J. Dijkema
- J. Leur, van de
- P.A. Zegeling
Abstract
In patients with acute heart failure, the heart can be assisted by the insertion of a mechanical device which takes over part of the heart’s work load by pumping blood from the left ventricle, one of the heart chambers, into the aorta. In this project, we formulate a model that describes the effect of such a device on the cardiovascular dynamics. We show that data for the pressure–volume relationship within a heart chamber that have been obtained experimentally can be reproduced quite accurately by our model. Moreover, such experimental data can help in calibrating unknown parameters that specify the characteristics of the pump. A key parameter turned out to be the extra friction that is encountered by the blood flowing through the heart pump.