Increased speed of analysis in directly coupled gas chromatography-mass spectrometry systems. II. Advantages of vacuum outlet operation of thick-film capillary columns

  • P.A. Leclercq
    ,
  • G.J. Scherpenzeel
    ,
  • E.A.A. Vermeer
    ,
  • C.A.M.G. Cramers
Output:
Contribution to journal
Article
Academic
Peer-review

Open access

Information

Output type

Output:
Contribution to journal
Article
Academic
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 61-71 (11 pages)

Journal (Volume, Issue Number)

Journal of Chromatography (Volume 241, Issue 1)

Publication milestones

  • Published - 1982

Publication status

Published - 1982

ISSN

0021-9673

Abstract

It is shown that direct insertion of the end of a capillary column into the ion source of a mass spectrometer increases the speed of analysis compared with atmospheric outlet conditions. This applies to thin-film columns, as reported previously [C. A. Cramers, G.J. Scherpenzeel and P.A. Leclercq, J. Chromatogr., 203 (1981) 207], as well as to the thick-film capillaries treated in this paper. For thick-film columns th enegative effect of diffusion in the stationary liquid phase on the speed of analysis, via an increased minimal plate height and a decreased optimal gas velocity, is more than compensated for by the effects of vacuum outlet operation. The gas chromatographic-mass spectrometric (GC-MS) coupling of capillary columns directly inserted into the ion source posesses the favourable properties of full sample transfer, no catalysis or adsorption in an interface and increased speed of analysis. Thick-film columns offer in addition a considerably increased sample capacity. The resolution of low-boiling compounds is enhanced and adsorption on the column wall is masked by the thicker film. Further, thick-film columns are less demanding with respect to instrument specifications such as speed of sampling, time constants and noise levels of detection and registration systems. The conclusion is that the best GC-MS interface appears to be no interface at all but rather direct insertion of the column end into the ion source.

Documents and Links

Metrics

Metrics

Scopus
citations
35
Download statistics
Download count
240

PlumX, opens in new tab

Citation count
35
Captures
6