Cryogenic thermometry with a common diode: Type BAS16
- A.P. Rijpma,
- H.J.M. Brake, ter
Output:
Contribution to journal
Article
Academic
Peer-reviewInformation
Output type
Output:
Contribution to journal
Article
Academic
Peer-reviewOriginal language
EnglishPages from-to (Number of pages)
Pages 68-69 (2 pages)Journal (Volume, Issue Number)
Cryogenics (Volume 46, Issue 1)Publication milestones
- Published - 2006
Publication status
Published - 2006
ISSN
0011-2275Abstract
Cryogenic test expts. often require a large no. of temps. to be monitored. In order to reduce cost, we investigated the feasibility of low-cost common diodes. We chose the Philips BAS16 diode in a type SOT23 package. By means of Stycast 2850FT, these diodes were glued into alumina holders. In total, 20 sensors were assembled and tested. With an excitation current of 100 micro A and considering a temp. range from 50 K to room temp., a quadratic least-squares-fit was obtained: T = 486.1 - 363.7V - 38.3V2. All sensors agreed with this fit within an error of 1.3 K.
Documents and Links
Metrics
Metrics
Scopus
citations
13
SciVal
citations
13
Download statistics
Download count
3
SciVal
FWCI
0.93
SciVal
Author count
2
SciVal
Paper percentile
70
PlumX, opens in new tab
Captures
10
Citation count
13