Skip to search boxSkip to navigationSkip to main content

Characterization of a novel carbonized foam electrode for wearable bio-potential recording

  • Hongyu Chen
    ,
  • Yuting Zhao
    ,
  • Zhenning Mei
    ,
  • Yongfeng Mei
    ,
  • Sidarto Bambang Oetomo
    ,
  • Wei Chen
  • Fudan University
    ,
  • Máxima Medical Center
    ,
  • Shanghai Key Laboratory of Medical Imaging Computing and Computer Assisted Intervention
Output:
Chapter in Book/Report/Conference proceeding
Conference contribution
Academic
Peer-review

Information

Output type

Output:
Chapter in Book/Report/Conference proceeding
Conference contribution
Academic
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 173-176 (4 pages)

Publication milestones

  • Published - 02/04/2018

Publication status

Published - 02/04/2018

Place of publication

Piscataway

Publisher

Institute of Electrical and Electronics Engineers, United States
978-1-5386-1110-4

ISBN (Electronic)

978-1-5386-1109-8

Host publication title

2018 IEEE 15th International Conference on Wearable and Implantable Body Sensor Networks, BSN 2018

Related Event

Title

15th International Conference on Wearable and Implantable Body Sensor Networks (BSN 2018)

Event type

Conference

Date

04/03/2018 - 07/03/2018

Location

Treasure Island Hotel & CasinoLas VegasUnited States

Abstract

<p>A novel dry disposable electrode using carbonized foam as conductive material is presented. The conductive material is flexible and the manufacturing of it is inexpensive. In this paper, the preparation of the conductive material and the electrical properties of the electrode are investigated. A test protocol is designed to compare the in-vitro impedance, skin-electrode interface and the signal quality of the proposed electrode with that of the wet Ag/AgCl electrode. Experimental results reveal that the carbonized foam has good flexibility and conductivity. The proposed electrode can acquire ECG signal of promising signal quality when compared with Ag/AgCl electrode in the case of static and motion. Furthermore, raw data with less power line interference was observed by proposed electrodes without noise suppression circuits or algorithms. All these make the novel electrode a promising candidate for wearable bio-potential recording.</p>

Metrics

Metrics

SciVal
citations
3
Scopus
citations
3
SciVal
FWCI
0.97
SciVal
Author count
6
SciVal
Paper percentile
70

PlumX, opens in new tab

Captures
10
Citation count
3

UN SDGs

In 2015, UN member states agreed to 17 global Sustainable Development Goals (SDGs) to end poverty, protect the planet and ensure prosperity for all. This output contributes to the following SDG(s):
  • SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well