Wireless Power Transfer to a Visual Prosthesis: 100 mW at 6.78 MHz

Tom van Nunen, Rob Mestrom, Hubregt Visser

Onderzoeksoutput: Hoofdstuk in Boek/Rapport/CongresprocedureConferentiebijdrageAcademicpeer review

3 Citaten (Scopus)
160 Downloads (Pure)

Samenvatting

An inductive wireless power transfer system for a neural implant is presented. The prototype can transfer 100 mW at 6.78 MHz transcutaneously for 0:08 ≤ k ≤ 0:31, which corresponds to a distance of 8 to 15 mm, and a lateral misalignment of 0 to 15 mm. The achieved efficiency ranges from 55 to 77 %, depending on the relative distance between the coils. The transmitter coil outer diameter is 55 mm, the receiver coil outside diameter is 35 mm. The inverter is class-E, using a GaN MOSFET, and the receiver is class-DE.
Originele taal-2Engels
Titel2021 IEEE International Symposium on Antennas and Propagation and North American Radio Science Meeting, APS/URSI 2021 - Proceedings
UitgeverijInstitute of Electrical and Electronics Engineers
Pagina's269-270
Aantal pagina's2
ISBN van elektronische versie978-1-7281-4670-6
DOI's
StatusGepubliceerd - 16 feb. 2022
Evenement2021 IEEE International Symposium on Antennas and Propagation and USNC/URSI National Radio Science Meeting, APSURSI 2021 - Marina Bay Sands, Singapore, Singapore
Duur: 4 dec. 202110 dec. 2021
https://2021apsursi.org/

Congres

Congres2021 IEEE International Symposium on Antennas and Propagation and USNC/URSI National Radio Science Meeting, APSURSI 2021
Verkorte titelAPSURSI 2021
Land/RegioSingapore
StadSingapore
Periode4/12/2110/12/21