The dual-mode dipole: A new array element for 7T body imaging with reduced SAR

Georgiy Solomakha, Carel van Leeuwen, Alexander Raaijmakers, Constantin Simovski, Alexander Popugaev, Redha Abdeddaim, Irina Melchakova, Stanislav Glybovski

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

Uittreksel

Purpose: To design and test an RF-coil based on two orthogonal eigenmodes in a pair of coupled dipoles, for 7 Tesla body imaging with improved SAR, called dual-mode dipole. Methods: The proposed coil consists of two dipoles and creates two orthogonal field distributions in a sample (the even and odd modes). A coupler used to excite the modes was miniaturized with the conductor track routing technique. Numerical simulations of the dual-mode dipole in the presence of a homogeneous phantom were performed. Moreover, an array of such coils was simulated with a voxel body model. For comparison, a fractionated dipole combined with a surface loop coil was also simulated. Both coils were tested in a 7 Tesla MRI system on a phantom. Subsequently four dual-mode dipoles or dipole/loop combinations were used for a comparison of imaging performance in a human volunteer. Results: Using the even mode of the dual-mode dipole showed 70% SAR reduction in comparison to the fractionated dipole while having the same (Formula presented.) in the prostate region. The odd mode of the dual-mode dipole showed a performance comparable to the surface loop both for SAR and B1 efficiency. The obtained results showed that the proposed coil while creating lower SAR gave images of the same quality as the reference coil. Conclusions: It was demonstrated that the array of dual-mode dipoles provided the same SNR and prostate imaging quality as the reference array, while demonstrating lower SAR. This is due to a smoother current distribution over a sample surface.

TaalEngels
Pagina's1459-1469
TijdschriftMagnetic Resonance in Medicine
Volume81
Nummer van het tijdschrift2
DOI's
StatusGepubliceerd - 1 feb 2019

Vingerafdruk

Prostate
Volunteers

Trefwoorden

    Citeer dit

    Solomakha, G., Leeuwen, C. V., Raaijmakers, A., Simovski, C., Popugaev, A., Abdeddaim, R., ... Glybovski, S. (2019). The dual-mode dipole: A new array element for 7T body imaging with reduced SAR. Magnetic Resonance in Medicine, 81(2), 1459-1469. DOI: 10.1002/mrm.27485
    Solomakha, Georgiy ; Leeuwen, Carel van ; Raaijmakers, Alexander ; Simovski, Constantin ; Popugaev, Alexander ; Abdeddaim, Redha ; Melchakova, Irina ; Glybovski, Stanislav. / The dual-mode dipole : A new array element for 7T body imaging with reduced SAR. In: Magnetic Resonance in Medicine. 2019 ; Vol. 81, Nr. 2. blz. 1459-1469
    @article{392449fe87ca443e8ddc92b6a20a12df,
    title = "The dual-mode dipole: A new array element for 7T body imaging with reduced SAR",
    abstract = "Purpose: To design and test an RF-coil based on two orthogonal eigenmodes in a pair of coupled dipoles, for 7 Tesla body imaging with improved SAR, called dual-mode dipole. Methods: The proposed coil consists of two dipoles and creates two orthogonal field distributions in a sample (the even and odd modes). A coupler used to excite the modes was miniaturized with the conductor track routing technique. Numerical simulations of the dual-mode dipole in the presence of a homogeneous phantom were performed. Moreover, an array of such coils was simulated with a voxel body model. For comparison, a fractionated dipole combined with a surface loop coil was also simulated. Both coils were tested in a 7 Tesla MRI system on a phantom. Subsequently four dual-mode dipoles or dipole/loop combinations were used for a comparison of imaging performance in a human volunteer. Results: Using the even mode of the dual-mode dipole showed 70{\%} SAR reduction in comparison to the fractionated dipole while having the same (Formula presented.) in the prostate region. The odd mode of the dual-mode dipole showed a performance comparable to the surface loop both for SAR and B1 efficiency. The obtained results showed that the proposed coil while creating lower SAR gave images of the same quality as the reference coil. Conclusions: It was demonstrated that the array of dual-mode dipoles provided the same SNR and prostate imaging quality as the reference array, while demonstrating lower SAR. This is due to a smoother current distribution over a sample surface.",
    keywords = "coil, Dipole, modes, parallel transmit, SAR",
    author = "Georgiy Solomakha and Leeuwen, {Carel van} and Alexander Raaijmakers and Constantin Simovski and Alexander Popugaev and Redha Abdeddaim and Irina Melchakova and Stanislav Glybovski",
    year = "2019",
    month = "2",
    day = "1",
    doi = "10.1002/mrm.27485",
    language = "English",
    volume = "81",
    pages = "1459--1469",
    journal = "Magnetic Resonance in Medicine",
    issn = "0740-3194",
    publisher = "Wiley",
    number = "2",

    }

    Solomakha, G, Leeuwen, CV, Raaijmakers, A, Simovski, C, Popugaev, A, Abdeddaim, R, Melchakova, I & Glybovski, S 2019, 'The dual-mode dipole: A new array element for 7T body imaging with reduced SAR' Magnetic Resonance in Medicine, vol. 81, nr. 2, blz. 1459-1469. DOI: 10.1002/mrm.27485

    The dual-mode dipole : A new array element for 7T body imaging with reduced SAR. / Solomakha, Georgiy; Leeuwen, Carel van; Raaijmakers, Alexander; Simovski, Constantin; Popugaev, Alexander; Abdeddaim, Redha; Melchakova, Irina; Glybovski, Stanislav.

    In: Magnetic Resonance in Medicine, Vol. 81, Nr. 2, 01.02.2019, blz. 1459-1469.

    Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

    TY - JOUR

    T1 - The dual-mode dipole

    T2 - Magnetic Resonance in Medicine

    AU - Solomakha,Georgiy

    AU - Leeuwen,Carel van

    AU - Raaijmakers,Alexander

    AU - Simovski,Constantin

    AU - Popugaev,Alexander

    AU - Abdeddaim,Redha

    AU - Melchakova,Irina

    AU - Glybovski,Stanislav

    PY - 2019/2/1

    Y1 - 2019/2/1

    N2 - Purpose: To design and test an RF-coil based on two orthogonal eigenmodes in a pair of coupled dipoles, for 7 Tesla body imaging with improved SAR, called dual-mode dipole. Methods: The proposed coil consists of two dipoles and creates two orthogonal field distributions in a sample (the even and odd modes). A coupler used to excite the modes was miniaturized with the conductor track routing technique. Numerical simulations of the dual-mode dipole in the presence of a homogeneous phantom were performed. Moreover, an array of such coils was simulated with a voxel body model. For comparison, a fractionated dipole combined with a surface loop coil was also simulated. Both coils were tested in a 7 Tesla MRI system on a phantom. Subsequently four dual-mode dipoles or dipole/loop combinations were used for a comparison of imaging performance in a human volunteer. Results: Using the even mode of the dual-mode dipole showed 70% SAR reduction in comparison to the fractionated dipole while having the same (Formula presented.) in the prostate region. The odd mode of the dual-mode dipole showed a performance comparable to the surface loop both for SAR and B1 efficiency. The obtained results showed that the proposed coil while creating lower SAR gave images of the same quality as the reference coil. Conclusions: It was demonstrated that the array of dual-mode dipoles provided the same SNR and prostate imaging quality as the reference array, while demonstrating lower SAR. This is due to a smoother current distribution over a sample surface.

    AB - Purpose: To design and test an RF-coil based on two orthogonal eigenmodes in a pair of coupled dipoles, for 7 Tesla body imaging with improved SAR, called dual-mode dipole. Methods: The proposed coil consists of two dipoles and creates two orthogonal field distributions in a sample (the even and odd modes). A coupler used to excite the modes was miniaturized with the conductor track routing technique. Numerical simulations of the dual-mode dipole in the presence of a homogeneous phantom were performed. Moreover, an array of such coils was simulated with a voxel body model. For comparison, a fractionated dipole combined with a surface loop coil was also simulated. Both coils were tested in a 7 Tesla MRI system on a phantom. Subsequently four dual-mode dipoles or dipole/loop combinations were used for a comparison of imaging performance in a human volunteer. Results: Using the even mode of the dual-mode dipole showed 70% SAR reduction in comparison to the fractionated dipole while having the same (Formula presented.) in the prostate region. The odd mode of the dual-mode dipole showed a performance comparable to the surface loop both for SAR and B1 efficiency. The obtained results showed that the proposed coil while creating lower SAR gave images of the same quality as the reference coil. Conclusions: It was demonstrated that the array of dual-mode dipoles provided the same SNR and prostate imaging quality as the reference array, while demonstrating lower SAR. This is due to a smoother current distribution over a sample surface.

    KW - coil

    KW - Dipole

    KW - modes

    KW - parallel transmit

    KW - SAR

    UR - http://www.scopus.com/inward/record.url?scp=85052928264&partnerID=8YFLogxK

    U2 - 10.1002/mrm.27485

    DO - 10.1002/mrm.27485

    M3 - Article

    VL - 81

    SP - 1459

    EP - 1469

    JO - Magnetic Resonance in Medicine

    JF - Magnetic Resonance in Medicine

    SN - 0740-3194

    IS - 2

    ER -

    Solomakha G, Leeuwen CV, Raaijmakers A, Simovski C, Popugaev A, Abdeddaim R et al. The dual-mode dipole: A new array element for 7T body imaging with reduced SAR. Magnetic Resonance in Medicine. 2019 feb 1;81(2):1459-1469. Beschikbaar vanaf, DOI: 10.1002/mrm.27485