Samenvatting
A polemic is given to show that the solute retention equation for micellar electrokinetic capillary chromatog. can be used to derive the capacity factor in parallel-current open tubular liq. chromatog. as developed by K. Slais et al. (J. Chromatog. 1992, 605, 167; J. Microcol. Sep. 1993, 5, 63). Differences in flow ratio caused by inhomogeneity of the retentive film will not affect the calcd. capacity factor in contrast to the dependence on flow ratio of the reduced capacity factor derived by Slais et al
Originele taal-2 | Engels |
---|---|
Pagina's (van-tot) | 187-189 |
Tijdschrift | Journal of Microcolumn Separations |
Volume | 5 |
Nummer van het tijdschrift | 3 |
DOI's | |
Status | Gepubliceerd - 1993 |