Performance of hot-electron bolometers for infrared photometry and X-ray micro-calorimetry

H.F.C. Hoevers, A.C. Bento, M.P. Bruijn, M. Frericks, F.B. Kiewiet, W.A. Mels, P.A.J. Korte

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

3 Citaten (Scopus)

Samenvatting

The performance of voltage biased superconducting transition edge bolometers for infrared photometry and X-ray micro-calorimetry is described. Experimental results for the resistance–temperature curve of the superconducting phase thermometer, the current–voltage characteristic of the bolometer, and its current noise are presented and evaluated. Excess current noise turns out to be present when the bolometers are biased in the superconducting transition. The possible origin of this excess noise is shortly discussed. For the infrared bolometer an NEP of 2 at 270 mK is obtained. The X-ray micro-calorimeter resolution is 60 eVFWHM at 155 mK for 5.9 keV X-rays, a factor 2.6 above the thermodynamical limit.
Originele taal-2Engels
Pagina's (van-tot)247-251
Aantal pagina's5
TijdschriftNuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume436
Nummer van het tijdschrift1-2
DOI's
StatusGepubliceerd - 1999