Optimum Tolerance for Load Mismatch and Coil Misalignment in Resonant Inductive WPT

Onderzoeksoutput: Hoofdstuk in Boek/Rapport/CongresprocedureConferentiebijdrageAcademicpeer review

Samenvatting

For inductively wirelessly powered medical implants we want to maximize the received power and want to have a large tolerance with respect to receiver load impedance or coil misalignment. To accomplish the first, we should optimize the load resistance for maximum Power Conversion Efficiency (PCE) and then apply a larger load resistance. To accomplish the latter, we should optimize for a maximum PCE for a setup wherein the mutual inductance is larger than expected in practice.

Originele taal-2Engels
Titel2022 Wireless Power Week, WPW 2022
UitgeverijInstitute of Electrical and Electronics Engineers
Pagina's576-579
Aantal pagina's4
ISBN van elektronische versie978-1-6654-8445-9
DOI's
StatusGepubliceerd - 17 aug. 2022
Extern gepubliceerdJa
Evenement2022 Wireless Power Week, WPW 2022 - Bordeaux, Frankrijk
Duur: 5 jul. 20228 jul. 2022

Congres

Congres2022 Wireless Power Week, WPW 2022
Land/RegioFrankrijk
StadBordeaux
Periode5/07/228/07/22

Vingerafdruk

Duik in de onderzoeksthema's van 'Optimum Tolerance for Load Mismatch and Coil Misalignment in Resonant Inductive WPT'. Samen vormen ze een unieke vingerafdruk.

Citeer dit