Optimal placement of bipolar surface EMG electrodes in the face based on single motor unit analysis

Bernd G Lapatki, Robert Oostenveld, Johannes P. van Dijk, Irmtrud E Jonas, Machiel J Zwarts, Dick F Stegeman

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

21 Citaten (Scopus)

Samenvatting

Locations of surface electromyography (sEMG) electrodes in the face are usually chosen on a macro-anatomical basis. In this study we describe optimal placement of bipolar electrodes based on a novel method and present results for lower facial muscles. We performed high-density sEMG recordings in 13 healthy participants. Raw sEMG signals were decomposed into motor unit action potentials (MUAPs). We positioned virtual electrode pairs in the interpolated monopolar MUAPs at different positions along muscle fiber direction and calculated the bipolar potentials. Electrode sites were determined where maximal bipolar amplitude was achieved and were validated. Objective guidelines for sEMG electrode placement improve the signal-to-noise ratio and may contribute to reduce cross talk, which is particularly important in the face. The method may be regarded as an important basis for improving the validity and reproducibility of sEMG in complex muscle areas.

Originele taal-2Engels
Pagina's (van-tot)299-314
Aantal pagina's16
TijdschriftPsychophysiology
Volume47
Nummer van het tijdschrift2
DOI's
StatusGepubliceerd - 1 mrt 2010
Extern gepubliceerdJa

Vingerafdruk Duik in de onderzoeksthema's van 'Optimal placement of bipolar surface EMG electrodes in the face based on single motor unit analysis'. Samen vormen ze een unieke vingerafdruk.

  • Citeer dit