Non-invasive blood oxygen saturation monitoring for neonates using reflectance pulse oximeter

W. Chen, I.B.I. Ayoola, S. Bambang Oetomo, L.M.G. Feijs

    Onderzoeksoutput: Hoofdstuk in Boek/Rapport/CongresprocedureConferentiebijdrageAcademicpeer review

    25 Citaten (Scopus)
    1 Downloads (Pure)

    Samenvatting

    Blood oxygen saturation is one of the key parameters for health monitoring of premature infants at the neonatal intensive care unit (NICU). In this paper, we propose and demonstrate a design of a wearable wireless blood saturation monitoring system. Reflectance pulse oxymeter based on Near Infrared Spectroscopy (NIRS) techniques are applied for enhancing the flexibility of measurements at different locations on the body of the neonates and the compatibility to be integrated into a non-invasive monitoring platform, such as a neonatal smart jacket. Prototypes with the reflectance sensors embedded in soft fabrics are built. The thickness of device is minimized to optimize comfort. To evaluate the performance of the prototype, experiments on the premature babies were carried out at NICU of Máxima Medical Centre (MMC) in Veldhoven, the Netherlands. The results show that the heart rate and SpO2 measured by the proposed design are corresponding to the readings of the standard monitor.
    Originele taal-2Engels
    TitelProceedings of the 2010 Design, Automation & Test in Europe Conference & Exhibition (DATE), 8-12 march 2010, Dresden, Germany
    Plaats van productiePiscataway
    UitgeverijInstitute of Electrical and Electronics Engineers
    Pagina's1530-1535
    ISBN van geprinte versie978-3-9810801-6-2
    DOI's
    StatusGepubliceerd - 2010
    Evenement13th Design, Automation and Test in Europe Conference and Exhibition (DATE 2010) - ICC, Dresden, Duitsland
    Duur: 8 mrt 201012 mrt 2010
    Congresnummer: 13
    https://www.date-conference.com/date10/

    Congres

    Congres13th Design, Automation and Test in Europe Conference and Exhibition (DATE 2010)
    Verkorte titelDATE 2010
    LandDuitsland
    StadDresden
    Periode8/03/1012/03/10
    Ander2010 EDAA I ACM SIGDA PhD Forum at DATE
    Internet adres

    Vingerafdruk Duik in de onderzoeksthema's van 'Non-invasive blood oxygen saturation monitoring for neonates using reflectance pulse oximeter'. Samen vormen ze een unieke vingerafdruk.

    Citeer dit