It is all in the wrist: wearable sleep staging in a clinical population versus reference polysomnography

Bernice Wulterkens (Corresponding author), Pedro Fonseca, Lieke W.A. Hermans, Marco Ross, Andreas Cerny, Peter Anderer, Xi Long, Johannes van Dijk, Nele L.E. Vandenbussche, Sigrid Pillen, Merel M. van Gilst, Sebastiaan Overeem

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

41 Citaten (Scopus)
75 Downloads (Pure)

Samenvatting

Purpose: There is great interest in unobtrusive long-term sleep measurements using wearable devices based on reflective photoplethysmography (PPG). Unfortunately, consumer devices are not validated in patient populations and therefore not suitable for clinical use. Several sleep staging algorithms have been developed and validated based on ECG-signals. However, translation from these techniques to data derived by wearable PPG is not trivial, and requires the differences between sensing modalities to be integrated in the algorithm, or having the model trained directly with data obtained with the target sensor. Either way, validation of PPG-based sleep staging algorithms requires a large dataset containing both gold standard measurements and PPG-sensor in the applicable clinical population. Here, we take these important steps towards unobtrusive, long-term sleep monitoring. Methods: We developed and trained an algorithm based on wrist-worn PPG and accelerometry. The method was validated against reference polysomnography in an independent clinical population comprising 244 adults and 48 children (age: 3 to 82 years) with a wide variety of sleep disorders. Results: The classifier achieved substantial agreement on four-class sleep staging with an average Cohen’s kappa of 0.62 and accuracy of 76.4%. For children/adolescents, it achieved even higher agreement with an average kappa of 0.66 and accuracy of 77.9%. Performance was significantly higher in non-REM parasomnias (kappa = 0.69, accuracy = 80.1%) and significantly lower in REM parasomnias (kappa = 0.55, accuracy = 72.3%). A weak correlation was found between age and kappa (ρ = −0.30, p<0.001) and age and accuracy (ρ = −0.22, p<0.001). Conclusion: This study shows the feasibility of automatic wearable sleep staging in patients with a broad variety of sleep disorders and a wide age range. Results demonstrate the potential for ambulatory long-term monitoring of clinical populations, which may improve diagnosis, estimation of severity and follow up in both sleep medicine and research.

Originele taal-2Engels
Pagina's (van-tot)885-897
Aantal pagina's13
TijdschriftNature and Science of Sleep
Volume13
DOI's
StatusGepubliceerd - 28 jun. 2021

Vingerafdruk

Duik in de onderzoeksthema's van 'It is all in the wrist: wearable sleep staging in a clinical population versus reference polysomnography'. Samen vormen ze een unieke vingerafdruk.

Citeer dit