Intramedullary fixation with screwed, conical stems : unsolicited results from animal experiments

P.J.M. Loon, van, H. Weinans, H.W.J. Huiskes, K. Groot, de, T.J.J.H. Slooff

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

2 Citaten (Scopus)
123 Downloads (Pure)

Samenvatting

For the purpose of studying bone remodeling around prostheses, a segmental replacement for the goat tibia was designed, using a conical, screw-threaded, hydroxyapatite-coated stem for fixation. Eight goats were provided with the implant, seven of which loosened within 10 days post-operatively, displaying progressive radiolucency and gross rotational motion. The eighth one also loosened radiographically, but developed a stabilizing callus bridge to prevent motion. A second design of similar shape and coating, but lacking the screw threads, was designed and also applied in eight animals. In this case, no loosening occurred in the first 6 weeks post-operatively. It is concluded that the application of screwed intramedullary stems for prosthetic fixation is not a viable concept, because the threads prevent the stem from subsiding and restabilizing when minor initial interface stress-relaxation and remodeling has occurred.
Originele taal-2Engels
Pagina's (van-tot)239-242
TijdschriftClinical Materials
Volume10
Nummer van het tijdschrift4
DOI's
StatusGepubliceerd - 1992

Vingerafdruk

Duik in de onderzoeksthema's van 'Intramedullary fixation with screwed, conical stems : unsolicited results from animal experiments'. Samen vormen ze een unieke vingerafdruk.

Citeer dit