Image-guided system versus manual marking for toric intraocular lens alignment in cataract surgery

V.S.C. Webers, N.J.C. Bauer, N. Visser, T.T.J.M. Berendschot, F.J.H.M. van den Biggelaar, R.M.M.A. Nuijts

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

59 Citaten (Scopus)
247 Downloads (Pure)

Samenvatting

Purpose To compare the accuracy of toric intraocular lens (IOL) alignment using the Verion Image-Guided System versus a conventional manual ink-marking procedure. Setting University Eye Clinic Maastricht, Maastricht, the Netherlands. Design Prospective randomized clinical trial. Methods Eyes with regular corneal astigmatism of at least 1.25 diopters (D) that required cataract surgery and toric IOL implantation (Acrysof SN6AT3-T9) were randomly assigned to the image-guided group or the manual-marking group. The primary outcome was the alignment of the toric IOL based on preoperative images and images taken immediately after surgery. Secondary outcome measures were residual astigmatism, uncorrected distance visual acuity (UDVA), and complications. Results The study enrolled 36 eyes (24 patients). The mean toric IOL misalignment was significantly less in the image-guided group than in the manual group 1 hour (1.3 degrees ± 1.6 [SD] versus 2.8 ± 1.8 degrees; P =.02) and 3 months (1.7 ± 1.5 degrees versus 3.1 ± 2.1 degrees; P <.05) postoperatively. The mean residual refractive cylinder was −0.36 ± 0.32 D and −0.47 ± 0.28 D in the image-guided group and manual group, respectively (P >.05). The mean UDVA was 0.03 ± 0.10 logarithm of minimum angle of resolution (logMAR) and 0.04 ± 0.09 logMAR, respectively (both P >.05). No intraoperative complications occurred during any surgery. Conclusion The IOL misalignment was significantly less with digital marking than with manual marking; this did not result in a better UDVA or lower residual refractive astigmatism.

Originele taal-2Engels
Pagina's (van-tot)781-788
Aantal pagina's8
TijdschriftJournal of Cataract and Refractive Surgery
Volume43
Nummer van het tijdschrift6
DOI's
StatusGepubliceerd - 1 jun. 2017
Evenement34 Congress of the European Society of Cataract and Refractive Surgeonsescrs 2016 - Bella Center, Copenhagen, Denemarken
Duur: 10 sep. 201614 sep. 2016
http://escrs.org/Copenhagen2016/default.asp

Vingerafdruk

Duik in de onderzoeksthema's van 'Image-guided system versus manual marking for toric intraocular lens alignment in cataract surgery'. Samen vormen ze een unieke vingerafdruk.

Citeer dit