Fluorescence detector cell for use in an integrated electrically driven separation system

P.K. Bokx, de, E.E.A. Gillissen, P. Weijer, van de, M.H.J. Bekkers, C.H.M. Bommel, van, J.G.M. Janssen

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

4 Citaten (Scopus)
164 Downloads (Pure)

Samenvatting

An integrated design for electrically driven separations is presented. The injector, column and detector are all located within the same cartridge, allowing for integral thermostating and short column lengths. Results on the performance of the proposed fluorescence detector are reported. Molar amounts as low as 2·10-19 mol of fluorescein (100-pl detector cell volume) can be detected. Short column lengths permit very fast separations, as is demonstrated by the separation of four laser dyes within 35 s.
Originele taal-2Engels
Pagina's (van-tot)115-121
TijdschriftJournal of Chromatography
Volume598
Nummer van het tijdschrift1
DOI's
StatusGepubliceerd - 1992

Vingerafdruk

Duik in de onderzoeksthema's van 'Fluorescence detector cell for use in an integrated electrically driven separation system'. Samen vormen ze een unieke vingerafdruk.

Citeer dit