Detection rate of fetal distress using contraction-dependent fetal heart rate variability analysis

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

Uittreksel

Objective: Monitoring of the fetal condition during labor is currently performed by cardiotocograpy (CTG). Despite the use of CTG in clinical practice, CTG interpretation suffers from a high inter- and intra-observer variability and a low specificity. In addition to CTG, analysis of fetal heart rate variability (HRV) has been shown to provide information on fetal distress. However, fetal HRV can be strongly influenced by uterine contractions, particularly during the second stage of labor. Therefore, the aim of this study is to examine if distinguishing contractions from rest periods can improve the detection rate of HRV features for fetal distress during the second stage of labor. Approach: We used a dataset of 100 recordings, containing 20 cases of fetuses with adverse outcome. The most informative HRV features were selected by a genetic algorithm and classification performance was evaluated using support vector machines. Main results: Classification performance of fetal heart rate segments closest to birth improved from a geometric mean of 70% to 79%. If the classifier was used to indicate fetal distress over time, the geometric mean at 15 minutes before birth improved from 60% to 72%. Significance: Our results show that combining contraction-dependent HRV features with HRV features calculated over the entire fetal heart rate signal improves the detection rate of fetal distress.

TaalEngels
Artikelnummer025008
Aantal pagina's12
TijdschriftPhysiological Measurement
Volume39
Nummer van het tijdschrift2
Vroegere onlinedatum19 jan 2018
DOI's
StatusGepubliceerd - 28 feb 2018

Vingerafdruk

Fetal Distress
Fetal Heart Rate
Heart Rate
Second Labor Stage
Parturition
Fetal Monitoring
Uterine Contraction
Observer Variation
Personnel
Fetus
Support vector machines
Classifiers
Genetic algorithms
Monitoring

Trefwoorden

    Citeer dit

    @article{14750236eac743568c4f59dbe429badc,
    title = "Detection rate of fetal distress using contraction-dependent fetal heart rate variability analysis",
    abstract = "Objective: Monitoring of the fetal condition during labor is currently performed by cardiotocograpy (CTG). Despite the use of CTG in clinical practice, CTG interpretation suffers from a high inter- and intra-observer variability and a low specificity. In addition to CTG, analysis of fetal heart rate variability (HRV) has been shown to provide information on fetal distress. However, fetal HRV can be strongly influenced by uterine contractions, particularly during the second stage of labor. Therefore, the aim of this study is to examine if distinguishing contractions from rest periods can improve the detection rate of HRV features for fetal distress during the second stage of labor. Approach: We used a dataset of 100 recordings, containing 20 cases of fetuses with adverse outcome. The most informative HRV features were selected by a genetic algorithm and classification performance was evaluated using support vector machines. Main results: Classification performance of fetal heart rate segments closest to birth improved from a geometric mean of 70{\%} to 79{\%}. If the classifier was used to indicate fetal distress over time, the geometric mean at 15 minutes before birth improved from 60{\%} to 72{\%}. Significance: Our results show that combining contraction-dependent HRV features with HRV features calculated over the entire fetal heart rate signal improves the detection rate of fetal distress.",
    keywords = "fetal heart rate variability, genetic algorithm, intrapartum monitoring, support vector machine, uterine contractions",
    author = "G.J.J. Warmerdam and R. Vullings and {van Laar}, J.O.E.H. and {van der Hout-van der Jagt}, M.B. and J.W.M. Bergmans and L. Schmitt and S.G. Oei",
    note = "{\circledC} 2018 Institute of Physics and Engineering in Medicine.",
    year = "2018",
    month = "2",
    day = "28",
    doi = "10.1088/1361-6579/aaa925",
    language = "English",
    volume = "39",
    journal = "Physiological Measurement",
    issn = "0967-3334",
    publisher = "Institute of Physics",
    number = "2",

    }

    Detection rate of fetal distress using contraction-dependent fetal heart rate variability analysis. / Warmerdam, G.J.J.; Vullings, R.; van Laar, J.O.E.H.; van der Hout-van der Jagt, M.B.; Bergmans, J.W.M.; Schmitt, L.; Oei, S.G.

    In: Physiological Measurement, Vol. 39, Nr. 2, 025008, 28.02.2018.

    Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

    TY - JOUR

    T1 - Detection rate of fetal distress using contraction-dependent fetal heart rate variability analysis

    AU - Warmerdam,G.J.J.

    AU - Vullings,R.

    AU - van Laar,J.O.E.H.

    AU - van der Hout-van der Jagt,M.B.

    AU - Bergmans,J.W.M.

    AU - Schmitt,L.

    AU - Oei,S.G.

    N1 - © 2018 Institute of Physics and Engineering in Medicine.

    PY - 2018/2/28

    Y1 - 2018/2/28

    N2 - Objective: Monitoring of the fetal condition during labor is currently performed by cardiotocograpy (CTG). Despite the use of CTG in clinical practice, CTG interpretation suffers from a high inter- and intra-observer variability and a low specificity. In addition to CTG, analysis of fetal heart rate variability (HRV) has been shown to provide information on fetal distress. However, fetal HRV can be strongly influenced by uterine contractions, particularly during the second stage of labor. Therefore, the aim of this study is to examine if distinguishing contractions from rest periods can improve the detection rate of HRV features for fetal distress during the second stage of labor. Approach: We used a dataset of 100 recordings, containing 20 cases of fetuses with adverse outcome. The most informative HRV features were selected by a genetic algorithm and classification performance was evaluated using support vector machines. Main results: Classification performance of fetal heart rate segments closest to birth improved from a geometric mean of 70% to 79%. If the classifier was used to indicate fetal distress over time, the geometric mean at 15 minutes before birth improved from 60% to 72%. Significance: Our results show that combining contraction-dependent HRV features with HRV features calculated over the entire fetal heart rate signal improves the detection rate of fetal distress.

    AB - Objective: Monitoring of the fetal condition during labor is currently performed by cardiotocograpy (CTG). Despite the use of CTG in clinical practice, CTG interpretation suffers from a high inter- and intra-observer variability and a low specificity. In addition to CTG, analysis of fetal heart rate variability (HRV) has been shown to provide information on fetal distress. However, fetal HRV can be strongly influenced by uterine contractions, particularly during the second stage of labor. Therefore, the aim of this study is to examine if distinguishing contractions from rest periods can improve the detection rate of HRV features for fetal distress during the second stage of labor. Approach: We used a dataset of 100 recordings, containing 20 cases of fetuses with adverse outcome. The most informative HRV features were selected by a genetic algorithm and classification performance was evaluated using support vector machines. Main results: Classification performance of fetal heart rate segments closest to birth improved from a geometric mean of 70% to 79%. If the classifier was used to indicate fetal distress over time, the geometric mean at 15 minutes before birth improved from 60% to 72%. Significance: Our results show that combining contraction-dependent HRV features with HRV features calculated over the entire fetal heart rate signal improves the detection rate of fetal distress.

    KW - fetal heart rate variability

    KW - genetic algorithm

    KW - intrapartum monitoring

    KW - support vector machine

    KW - uterine contractions

    UR - http://www.scopus.com/inward/record.url?scp=85043519894&partnerID=8YFLogxK

    U2 - 10.1088/1361-6579/aaa925

    DO - 10.1088/1361-6579/aaa925

    M3 - Article

    VL - 39

    JO - Physiological Measurement

    T2 - Physiological Measurement

    JF - Physiological Measurement

    SN - 0967-3334

    IS - 2

    M1 - 025008

    ER -