TY - CONF
T1 - Design of a compact robotic assisted ophthalmic system
AU - Meenink, H.C.M.
AU - Steinbuch, M.
AU - Rosielle, P.C.J.N.
N1 - Proceedings of the Annual Meeting of the Association for Research in Vision and Ophthalmology, 30 April - 6 May 2011, Fort Lauderdale, Florida
PY - 2011
Y1 - 2011
N2 - Purpose:Robotics have enhanced and refined microinvasive surgery in several disciplines. Its applicability in eye surgery has been limited by ergonomic and scaling issues. Our aim was to design and build a microrobotic system adapted to the needs of vitreoretinal surgeons.
Methods:Constraints regarding head positioning and size, ocular access, surgical execution, and procedural requirements were defined by observations at live surgeries, discussions with surgeons, operation room teams, and computer simulations. Additional design parameters for the robotic slave (RS) included a low weight, high stiffness, low friction and play-free design. For the control module (CM), intuitiveness of the controller, body posture of the operator and patient proximity were considered.
Results:The RS consists of at least two instrument manipulators (IMs).The IM’s design allows 5 degrees of freedom through a kinematically defined rotation point at the entry site into the sclera. Force measurement down to 10mN is possible and manipulation with an accuracy of
AB - Purpose:Robotics have enhanced and refined microinvasive surgery in several disciplines. Its applicability in eye surgery has been limited by ergonomic and scaling issues. Our aim was to design and build a microrobotic system adapted to the needs of vitreoretinal surgeons.
Methods:Constraints regarding head positioning and size, ocular access, surgical execution, and procedural requirements were defined by observations at live surgeries, discussions with surgeons, operation room teams, and computer simulations. Additional design parameters for the robotic slave (RS) included a low weight, high stiffness, low friction and play-free design. For the control module (CM), intuitiveness of the controller, body posture of the operator and patient proximity were considered.
Results:The RS consists of at least two instrument manipulators (IMs).The IM’s design allows 5 degrees of freedom through a kinematically defined rotation point at the entry site into the sclera. Force measurement down to 10mN is possible and manipulation with an accuracy of
M3 - Poster
SP - E-Abstract 6125-
ER -