Samenvatting
There is an incorrect representation of the expression for resistances in parallel in equation (1) in section 4.1 'Voltage probe calibration' on page 6. The numerator and denominator in the equation are reversed and should read: I = Uc Rm + Rt/RmRt. Rm is the measuring resistor, Rt the terminating resistor at the oscilloscope and Uc is the voltage drop across Rm induced by the current I. None of the calculations and conclusions of the paper are affected. The authors apologise for any confusion that this transcription error may have caused.
Originele taal-2 | Engels |
---|---|
Artikelnummer | 029503 |
Aantal pagina's | 2 |
Tijdschrift | Journal of Physics D: Applied Physics |
Volume | 52 |
Nummer van het tijdschrift | 2 |
DOI's |
|
Status | Gepubliceerd - 9 jan. 2019 |
Vingerafdruk
Duik in de onderzoeksthema's van 'Corrigendum: Concepts and characteristics of the 'COST Reference Microplasma Jet' (Journal of Physics D: Applied Physics (2016) 49 (084003) DOI: 10.1088/0022-3727/49/8/084003)'. Samen vormen ze een unieke vingerafdruk.Impact
-
Plasmas for biological, environmental and medical applications
Sobota, A. (Content manager)
Impact: Research Topic/Theme (at group level)