Comparative dosimetry of radiography device, MSCT device and two CBCT devices in the elbow region

Juha Koivisto (Corresponding author), Maureen van Eijnatten, John Ludlow, Timo Kiljunen, Xie-Qi Shi, Jan Wolff

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

Samenvatting

The aim of the study was to estimate and to compare effective doses in the elbow region resulting from four different x-ray imaging modalities. Absorbed organ doses were measured using 11 metal oxide field effect transistor (MOSFET) dosimeters that were placed in a custom-made anthropomorphic elbow RANDO phantom. Examinations were performed using Shimadzu FH-21 HR radiography device, Siemens Sensation Open 24-slice MSCT-device, NewTom 5G CBCT device, and Planmed Verity CBCT device, and the effective doses were calculated according to ICRP 103 recommendations. The effective dose for the conventional radiographic device was 1.5 µSv. The effective dose for the NewTom 5G CBCT ranged between 2.0 and 6.7 µSv, for the Planmed Verity CBCT device 2.6 µSv and for the Siemens Sensation MSCT device 37.4 µSv. Compared with conventional 2D radiography, this study demonstrated a 1.4-4.6 fold increase in effective dose for CBCT and 25-fold dose for standard MSCT protocols. When compared with 3D CBCT protocols, the study showed a 6-19 fold increase in effective dose using a standard MSCT protocol. CBCT devices offer a feasible low-dose alternative for elbow 3D imaging when compared to MSCT.

Originele taal-2Engels
Pagina's (van-tot)128-138
Aantal pagina's11
TijdschriftJournal of Applied Clinical Medical Physics
Volume22
Nummer van het tijdschrift5
Vroegere onlinedatum3 apr 2021
DOI's
StatusGepubliceerd - mei 2021

Bibliografische nota

© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Vingerafdruk Duik in de onderzoeksthema's van 'Comparative dosimetry of radiography device, MSCT device and two CBCT devices in the elbow region'. Samen vormen ze een unieke vingerafdruk.

Citeer dit