Clinical and radiologic outcomes of a fully hydroxyapatite-coated femoral revision stem: excessive stress shielding incidence and its consequences

Ilknur Sanli, J.J.C. Arts, Jan Geurts

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

9 Citaten (Scopus)

Samenvatting

Stress shielding remains a concern in total hip arthroplasty. The consequences of stress shielding in hydroxyapatite-coated femoral component revisions were evaluated in a prospective cohort study. A total of 106 patients operated on by revision total hip arthroplasty were identified. Sixty-three patients were eligible for clinical and radiologic assessment of osseointegration, bone remodeling, and stress shielding. Five patients showed evidence of excessive stress shielding. One patient experienced a periprosthetic fracture. No adverse events occurred in the remaining patients with a low rate of thigh pain and reliable osseointegration. This is the only available study concerning mid- to long-term consequences of excessive stress shielding in hydroxyapatite-coated revision stems. We advocate surgeons using these stems to remain vigilant and be aware of possible stress shielding side effects.

Originele taal-2Engels
Pagina's (van-tot)209-214
Aantal pagina's6
TijdschriftJournal of Arthroplasty
Volume31
Nummer van het tijdschrift1
DOI's
StatusGepubliceerd - 1 jan 2016