Uittreksel

Objective: To date, mainly obtrusive methods (e.g., adhesive electrodes in electroencephalography or electrocardiography) are necessary to determine the preterm infant sleep states. As any obtrusive measure should be avoided in preterm infants because of their immature skin development, we investigated the possibility of automated sleep staging using electrocardiograph signals from non-adhesive capacitive electrocardiography.
Approach: Capacitive electrocardiography data from eight different patients with a mean gestational age of 30 ± 2.5 weeks are compared to manually annotated reference signals from classic adhesive electrodes. The sleep annotations were performed by two trained observers based on behavioral observations.
Main results: Based on these annotations, classification performance of the preterm infant active and quiet sleep states, based on capacitive electrocardiography signals, showed a kappa value of 0.56 ± 0.20. Adding wake and caretaking into the classification, a performance of Kappa 0.44 ± 0.21 was achieved. In-between sleep state performance showed a classification performance of Kappa 0.36 ± 0.12. Lastly, a performance for all sleep states of Kappa 0.35 ± 0.17 was attained.
Significance: capacitive electrocardiography signals can be utilized to classify the central preterm infant sleep states, active and quiet sleep. With further research based on our results, automated classification of sleep states can become an essential instrument in the future intensive neonatal care for continuous brain maturation monitoring. Especially, being able to use capacitive electrocardiography for continuous monitoring is a significant contributor to reducing disruption and harm for this extreme fragile patient group.
Originele taal-2Engels
Artikelnummer055003
Aantal pagina's14
TijdschriftPhysiological Measurement
Volume40
Nummer van het tijdschrift5
Vroegere onlinedatum21 mrt 2019
DOI's
StatusGepubliceerd - 4 jun 2019

Vingerafdruk

Electrocardiography
Premature Infants
Sleep
Adhesives
Electrodes
Neonatal Intensive Care
Monitoring
Electroencephalography
Gestational Age
Brain
Skin
Research

Citeer dit

@article{fc942f622c6f4b30907ff37458426b48,
title = "Automated preterm infant sleep staging using capacitive electrocardiography",
abstract = "Objective: To date, mainly obtrusive methods (e.g., adhesive electrodes in electroencephalography or electrocardiography) are necessary to determine the preterm infant sleep states. As any obtrusive measure should be avoided in preterm infants because of their immature skin development, we investigated the possibility of automated sleep staging using electrocardiograph signals from non-adhesive capacitive electrocardiography. Approach: Capacitive electrocardiography data from eight different patients with a mean gestational age of 30 ± 2.5 weeks are compared to manually annotated reference signals from classic adhesive electrodes. The sleep annotations were performed by two trained observers based on behavioral observations. Main results: Based on these annotations, classification performance of the preterm infant active and quiet sleep states, based on capacitive electrocardiography signals, showed a kappa value of 0.56 ± 0.20. Adding wake and caretaking into the classification, a performance of Kappa 0.44 ± 0.21 was achieved. In-between sleep state performance showed a classification performance of Kappa 0.36 ± 0.12. Lastly, a performance for all sleep states of Kappa 0.35 ± 0.17 was attained.Significance: capacitive electrocardiography signals can be utilized to classify the central preterm infant sleep states, active and quiet sleep. With further research based on our results, automated classification of sleep states can become an essential instrument in the future intensive neonatal care for continuous brain maturation monitoring. Especially, being able to use capacitive electrocardiography for continuous monitoring is a significant contributor to reducing disruption and harm for this extreme fragile patient group.",
keywords = "automated sleep classification, capacitive ECG, classic ECG, machine learning, preterm infants, unobtrusive",
author = "Jan Werth and Aline Serteyn and Peter Andriessen and Ronald Aarts and Xi Long",
year = "2019",
month = "6",
day = "4",
doi = "10.1088/1361-6579/ab1224",
language = "English",
volume = "40",
journal = "Physiological Measurement",
issn = "0967-3334",
publisher = "Institute of Physics",
number = "5",

}

Automated preterm infant sleep staging using capacitive electrocardiography. / Werth, Jan (Corresponding author); Serteyn, Aline; Andriessen, Peter; Aarts, Ronald; Long, Xi.

In: Physiological Measurement, Vol. 40, Nr. 5, 055003, 04.06.2019.

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

TY - JOUR

T1 - Automated preterm infant sleep staging using capacitive electrocardiography

AU - Werth, Jan

AU - Serteyn, Aline

AU - Andriessen, Peter

AU - Aarts, Ronald

AU - Long, Xi

PY - 2019/6/4

Y1 - 2019/6/4

N2 - Objective: To date, mainly obtrusive methods (e.g., adhesive electrodes in electroencephalography or electrocardiography) are necessary to determine the preterm infant sleep states. As any obtrusive measure should be avoided in preterm infants because of their immature skin development, we investigated the possibility of automated sleep staging using electrocardiograph signals from non-adhesive capacitive electrocardiography. Approach: Capacitive electrocardiography data from eight different patients with a mean gestational age of 30 ± 2.5 weeks are compared to manually annotated reference signals from classic adhesive electrodes. The sleep annotations were performed by two trained observers based on behavioral observations. Main results: Based on these annotations, classification performance of the preterm infant active and quiet sleep states, based on capacitive electrocardiography signals, showed a kappa value of 0.56 ± 0.20. Adding wake and caretaking into the classification, a performance of Kappa 0.44 ± 0.21 was achieved. In-between sleep state performance showed a classification performance of Kappa 0.36 ± 0.12. Lastly, a performance for all sleep states of Kappa 0.35 ± 0.17 was attained.Significance: capacitive electrocardiography signals can be utilized to classify the central preterm infant sleep states, active and quiet sleep. With further research based on our results, automated classification of sleep states can become an essential instrument in the future intensive neonatal care for continuous brain maturation monitoring. Especially, being able to use capacitive electrocardiography for continuous monitoring is a significant contributor to reducing disruption and harm for this extreme fragile patient group.

AB - Objective: To date, mainly obtrusive methods (e.g., adhesive electrodes in electroencephalography or electrocardiography) are necessary to determine the preterm infant sleep states. As any obtrusive measure should be avoided in preterm infants because of their immature skin development, we investigated the possibility of automated sleep staging using electrocardiograph signals from non-adhesive capacitive electrocardiography. Approach: Capacitive electrocardiography data from eight different patients with a mean gestational age of 30 ± 2.5 weeks are compared to manually annotated reference signals from classic adhesive electrodes. The sleep annotations were performed by two trained observers based on behavioral observations. Main results: Based on these annotations, classification performance of the preterm infant active and quiet sleep states, based on capacitive electrocardiography signals, showed a kappa value of 0.56 ± 0.20. Adding wake and caretaking into the classification, a performance of Kappa 0.44 ± 0.21 was achieved. In-between sleep state performance showed a classification performance of Kappa 0.36 ± 0.12. Lastly, a performance for all sleep states of Kappa 0.35 ± 0.17 was attained.Significance: capacitive electrocardiography signals can be utilized to classify the central preterm infant sleep states, active and quiet sleep. With further research based on our results, automated classification of sleep states can become an essential instrument in the future intensive neonatal care for continuous brain maturation monitoring. Especially, being able to use capacitive electrocardiography for continuous monitoring is a significant contributor to reducing disruption and harm for this extreme fragile patient group.

KW - automated sleep classification

KW - capacitive ECG

KW - classic ECG

KW - machine learning

KW - preterm infants

KW - unobtrusive

UR - http://www.scopus.com/inward/record.url?scp=85067219864&partnerID=8YFLogxK

U2 - 10.1088/1361-6579/ab1224

DO - 10.1088/1361-6579/ab1224

M3 - Article

C2 - 30897551

AN - SCOPUS:85067219864

VL - 40

JO - Physiological Measurement

JF - Physiological Measurement

SN - 0967-3334

IS - 5

M1 - 055003

ER -