Arterial and portal venous liver perfusion using selective spin labelling MRI

  • H.J. Schalkx
  • , E.T. (Esben) Petersen
  • , N.H.G.M. (Nicky) Peters
  • , W.B. Veldhuis
  • , M.S. Leeuwen, van
  • , J.P.W. Pluim
  • , M.A.A.J. Bosch, van den
  • , M. Stralen, van

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

614 Downloads (Pure)

Samenvatting

Purpose
To investigate the feasibility of selective arterial and portal venous liver perfusion imaging with spin labelling (SL) MRI, allowing separate labelling of each blood supply.

Methods
The portal venous perfusion was assessed with a pulsed EPISTAR technique and the arterial perfusion with a pseudo-continuous sequence. To explore precision and reproducibility, portal venous and arterial perfusion were separately quantified in 12 healthy volunteers pre- and postprandially (before and after meal intake). In a subgroup of 6 volunteers, the accuracy of the absolute portal perfusion and its relative postprandial change were compared with MRI flow measurements of the portal vein.

Results
The portal venous perfusion significantly increased from 63 ± 22 ml/100g/min preprandially to 132 ± 42 ml/100g/min postprandially. The arterial perfusion was lower with 35 ± 22 preprandially and 22 ± 30 ml/100g/min postprandially. The pre- and postprandial portal perfusion using SL correlated well with flow-based perfusion (r2 = 0.71). Moreover, postprandial perfusion change correlated well between SL- and flow-based quantification (r2 = 0.77). The SL results are in range with literature values.

Conclusion
Selective spin labelling MRI of the portal venous and arterial blood supply successfully quantified liver perfusion. This non-invasive technique provides specific arterial and portal venous perfusion imaging and could benefit clinical settings where contrast agents are contraindicated.
Originele taal-2Engels
Pagina's (van-tot)1529-1540
Aantal pagina's12
TijdschriftEuropean Radiology
Volume25
Nummer van het tijdschrift6