An amperometric detector has been constructed for use with capillary LC columns. The micro glassy-carbon working electrode and the capillary column outlet form a micro wall-jet detector cell. This detector cell is immersed in a vessel containing an Ag/AgCl reference and a Pt counter electrode and has an effective volume between 1 and 0.1 nl. The operational parameters of
the detector were determined using a chromatographic system with various fused-silica capillary columns (5 to 30 mum i.d.). The detector has been found to match the performance of 26 mum i.d. capillary columns in terms of sensitivity and variance.
An example of application. the chromatographic determination of some phenols is presented.
| Originele taal-2 | Engels |
|---|---|
| Titel | Eighth International Symposium on Capillary Chromatography : Riva del Garda, Italy, May 19-21, 1987. |
| Redacteuren | P. Sandra |
| Plaats van productie | Heidelberg |
| Uitgeverij | Hüthig |
| Pagina's | 1122-1129 |
| Volume | 2 |
| Status | Gepubliceerd - 1987 |
| Evenement | 8th International Symposium on Capillary Chromatography, 1987 - Palazzo dei Congressi, Riva del Garda, Italië Duur: 19 mei 1987 → 21 mei 1987 |
Congres
| Congres | 8th International Symposium on Capillary Chromatography, 1987 |
|---|---|
| Land/Regio | Italië |
| Stad | Riva del Garda |
| Periode | 19/05/87 → 21/05/87 |
| Ander | Eighth International Symposium on Capillary Chromatography : Baltimore, May 25-29, 1992 |
Vingerafdruk
Duik in de onderzoeksthema's van 'An on-column amperometric detector for capillary liquid chomatography'. Samen vormen ze een unieke vingerafdruk.Citeer dit
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver