The authors describe an automated calorimeter for measurement of the sp. heats at 10 > T > 0.5 K. It uses sample of moderate size (100-1000 mg), has a moderate precision and accuracy (2%-5%), is easy to operate and the measurements can be done quickly with He4. The accuracy of this calorimeter was checked by measurement of sp. heat of copper and that of aluminum near its superconducting transition temp. [on SciFinder (R)]