TY - JOUR
T1 - A multi-aperture encoding scheme for increased SNR in photoacoustic Imaging
AU - Gholampour, Amir
AU - Cano, Camilo
AU - van Sambeek, Marc R.H.M.
AU - Lopata, Richard
AU - Wu, Min
AU - Schwab, Hans-Martin
N1 - Publisher Copyright:
© 2024
PY - 2024/6
Y1 - 2024/6
N2 - Photoacoustic imaging creates light-induced ultrasonic signals to provide valuable information on internal body structures and tissue morphology non-invasively. A multi-aperture photoacoustic imaging (MP-PAI) system is an improvement over conventional photoacoustic imaging (PAI) systems in terms of resolution, contrast, and field of view. Previously, a prototype MP-PAI system was introduced based on multiple capacitive micromachined ultrasound transducers (CMUTs) with shared channels, such that each element in a CMUT shares its channel with its counterpart in other CMUTs. The system uses the biasing voltages of the CMUTs to switch between them and multiplex the received signals in time. Notwithstanding all the enhancements, the signal-to-noise ratio (SNR) remains limited in PAI. To address this issue, we are proposing a multi-aperture encoding scheme (MAES) to further increase the SNR in a multi-aperture PAI system. The proposed method involves receiving signals with multiple CMUTs simultaneously based on an encoding matrix, instead of switching between individual CMUTs. As a result, shared channels contain a superposition of signals, which are later recovered by applying a decoding matrix. Here, an analytical model for computing SNR with an arbitrary encoding sequence is presented, and the method is validated through numerical simulations and in an experimental study. Bipolar and unipolar encoding sequences were considered for the experiments. The numerical results show, in comparison to conventional MP-PAI, that MAES will obtain an SNR gain of 5.8 and 8.8 dB for S-sequence and truncated Hadamard encodings, respectively, when using 15 transducers. In experiments, three transducers are encoded by S-sequences and show 1.5 dB improvement in SNR over conventional MP-PAI method, which aligns well with the analytical model.
AB - Photoacoustic imaging creates light-induced ultrasonic signals to provide valuable information on internal body structures and tissue morphology non-invasively. A multi-aperture photoacoustic imaging (MP-PAI) system is an improvement over conventional photoacoustic imaging (PAI) systems in terms of resolution, contrast, and field of view. Previously, a prototype MP-PAI system was introduced based on multiple capacitive micromachined ultrasound transducers (CMUTs) with shared channels, such that each element in a CMUT shares its channel with its counterpart in other CMUTs. The system uses the biasing voltages of the CMUTs to switch between them and multiplex the received signals in time. Notwithstanding all the enhancements, the signal-to-noise ratio (SNR) remains limited in PAI. To address this issue, we are proposing a multi-aperture encoding scheme (MAES) to further increase the SNR in a multi-aperture PAI system. The proposed method involves receiving signals with multiple CMUTs simultaneously based on an encoding matrix, instead of switching between individual CMUTs. As a result, shared channels contain a superposition of signals, which are later recovered by applying a decoding matrix. Here, an analytical model for computing SNR with an arbitrary encoding sequence is presented, and the method is validated through numerical simulations and in an experimental study. Bipolar and unipolar encoding sequences were considered for the experiments. The numerical results show, in comparison to conventional MP-PAI, that MAES will obtain an SNR gain of 5.8 and 8.8 dB for S-sequence and truncated Hadamard encodings, respectively, when using 15 transducers. In experiments, three transducers are encoded by S-sequences and show 1.5 dB improvement in SNR over conventional MP-PAI method, which aligns well with the analytical model.
KW - CMUT
KW - Coherent compounding
KW - Encoding
KW - Multi-aperture
KW - Photoacoustic imaging
KW - SNR
UR - http://www.scopus.com/inward/record.url?scp=85185893662&partnerID=8YFLogxK
U2 - 10.1016/j.pacs.2024.100598
DO - 10.1016/j.pacs.2024.100598
M3 - Article
C2 - 39670192
AN - SCOPUS:85185893662
SN - 2213-5979
VL - 37
JO - Photoacoustics
JF - Photoacoustics
M1 - 100598
ER -