A flexible five-channel shielded-coaxial-cable (SCC) transceive neck coil for high resolution carotid imaging at 7T

Thomas Ruytenberg, Andrew Webb, Irena Zivkovic (Corresponding author)

Onderzoeksoutput: Bijdrage aan tijdschriftTijdschriftartikelAcademicpeer review

Samenvatting

Purpose
Imaging the carotid arteries at 7T ideally requires a flexible multichannel array that allows B1‐shimming and conforms to different neck sizes. The major challenge is to minimize coupling between closely spaced coils and to make the coupling relatively insensitive to loading conditions.

Methods
We have designed a five‐channel flexible transceive array composed of shielded‐coaxial‐cable coils placed on the anterior part of the neck and conforming to the anatomy. In vivo imaging of the carotid arteries in three subjects has been performed.
Results
The measured noise correlation matrices show the decoupling level between the individual elements to be −12.5 dB and better. Anatomical localizer imaging of the carotids shows both carotids in every subject well visualized after B1‐shimming. In vivo black‐blood, carotid images were acquired with very high in‐plane spatial resolution (0.25 × 0.25 mm2) with clear depiction of the vessel walls.

Conclusions
The flexibility of the proposed coil has been demonstrated by imaging subjects with different neck circumferences. To the best of our knowledge, the in‐plane resolution of 0.25 × 0.25 mm2 is the highest reported at 7T.
Originele taal-2Engels
Pagina's (van-tot)1672-1677
Aantal pagina's6
TijdschriftMagnetic Resonance in Medicine
Volume84
Nummer van het tijdschrift3
Vroegere onlinedatum12 feb 2020
DOI's
StatusE-publicatie vóór gedrukte publicatie - 12 feb 2020
Extern gepubliceerdJa