TRIPLE Q : a three channel quantum cascade laser absorption spectrometer for fast multiple species concentration measurements

M. Hübner, S. Welzel, D. Marinov, O. Guaitella, S. Glitsch, A. Rousseau, J. Röpcke

Research output: Contribution to journalArticleAcademicpeer-review

22 Citations (Scopus)
112 Downloads (Pure)

Abstract

A compact and transportable three channel quantum cascade laser system (TRIPLE Q) based on mid-infrared absorption spectroscopy has been developed for time-resolved plasma diagnostics. The TRIPLE Q spectrometer encompasses three independently controlled quantum cascade lasers (QCLs), which can be used for chemical sensing, particularly for gas phase analysis of plasmas. All three QCLs are operated in the intra-pulse mode with typical pulse lengths of the order of 150 ns. Using a multiplexed detection, a time resolution shorter than 1 µs can be achieved. Hence, the spectrometer is well suited to study kinetic processes of multiple infrared active compounds in reactive plasmas. A special data processing and analysis technique has been established to account for time jitter effects of the infrared emission of the QCLs. The performance of the TRIPLE Q system has been validated in pulsed direct current plasmas containing N2O/air and NO2/air.
Original languageEnglish
Article number093102
Pages (from-to)1-7
Number of pages7
JournalReview of Scientific Instruments
Volume82
DOIs
Publication statusPublished - 2011

Fingerprint

Dive into the research topics of 'TRIPLE Q : a three channel quantum cascade laser absorption spectrometer for fast multiple species concentration measurements'. Together they form a unique fingerprint.

Cite this