Abstract

The shielded-coaxial-cable (SCC) coils have been proposed recently as highly decoupled elements per se. The SCC coils are not very sensitive to bending, elongation and various degrees of overlapping. We experimentally investigated the central SNR when the SCC coils are placed in three different arrangements on a cylindrical object. The highest central SNR (more than 30% higher than in the other two arrangements) was obtained when the coils were elongated and adjacent to each other. This arrangement is important when the ROI is located deep inside the field of view, such in c-spine spinal cord or deep brain imaging.
Original languageEnglish
Number of pages1
Publication statusPublished - 2022
EventJoint Annual Meeting ISMRM-ESMRMB & ISMRT 31st Annual Meeting - London, United Kingdom
Duration: 7 May 202212 May 2022
Conference number: 31

Conference

ConferenceJoint Annual Meeting ISMRM-ESMRMB & ISMRT 31st Annual Meeting
Country/TerritoryUnited Kingdom
CityLondon
Period7/05/2212/05/22

Fingerprint

Dive into the research topics of 'Shielded Coaxial Cable Coils: the array configuration for maximized central SNR at 7T MRI'. Together they form a unique fingerprint.

Cite this