Abstract
Background and purpose: It is estimated that one third of the institutes applying clinical beta sources does not perform
independent dosimetry. The Netherlands commission on radiation dosimetry (NCS) recently published recommended
quality control procedures and detectors for the dosimetry of beta sources. The main issues of NCS Report 14 are
summarized here.
Materials and methods: A dosimetry survey was performed among 23 institutes in The Netherlands and Belgium. Well
ionization chambers, a plastic scintillator, plane-parallel ionization chamber, diode and radiochromic film were used for
determination of source strength (dose rate at reference distance) and uniformity of intravascular and ophthalmic
sources. The source strength of multiple sources of each type was measured and compared with the source strength
specified by the manufacturer.
Results: The standard deviation of the difference between measured and specified source strength was mostly about
3%, but varied between 0.8 and 15.8% depending on factors such as source type, detector, phantom and manufacturers
calibration. The average non-uniformity was about 7% for intravascular sources and 20% for ophthalmic sources. It is
estimated that the total relative standard uncertainty can be kept below G4% (1s) with all detectors tested. Maximum
deviations in source strength of 10% and a non-uniformity below 10% (intravascular) and 30% (ophthalmic) are
recommended.
Conclusions: Dosimetric and non-dosimetric quality control procedures on beta sources are recommended. They enable
standardized measurements, including the determination of relative source strength and non-uniformity. Absolute
calibrations depend on the future introduction of primary standards for clinical beta sources.
Original language | English |
---|---|
Pages (from-to) | 223-229 |
Journal | Radiotherapy and Oncology |
Volume | 78 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2006 |