Reassembling and testing of a high-precision heat capacity drop calorimeter. Heat capacity of some polyphenyls at T = 298.15 K

L.M.N.B.F. Santos, M.A. Alves da Rocha, A.S.M.C. Rodrigues, V. Štejfa, M. Fulem, M. Bastos

    Research output: Contribution to journalArticleAcademicpeer-review

    32 Citations (Scopus)

    Abstract

    The description of the reassembling and testing of a twin heat conduction, high-precision, drop microcalorimeter for the measurement of heat capacities of small samples are presented. The apparatus, originally developed and used at the Thermochemistry Laboratory, Lund, Sweden, has now been reassembled and modernized, with changes being made as regarding temperature sensors, electronics and data acquisition system. The apparatus was thereafter thoroughly tested, using benzoic acid and hexafluorobenzene as test substances. The accuracy of the View the MathML sourceCp,m° (298.15 K) data obtained with this apparatus is comparable to that achieved by high-precision adiabatic calorimetry. Here we also present the results of heat capacity measurements on of some polyphenyls (1,2,3-triphenylbenzene, 1,3,5-triphenylbenzene, p-terphenyl, m-terphenyl, o-terphenyl, p-quaterphenyl) at T = 298.15 K, measured with the renewed high precision heat capacity drop calorimeter system. The high resolution and accuracy of the obtained heat capacity data enabled differentiation among the ortho-, meta-, and para-phenyl isomers.
    Original languageEnglish
    Pages (from-to)1818-1823
    JournalThe Journal of Chemical Thermodynamics
    Volume23
    Issue number12
    DOIs
    Publication statusPublished - 2011

    Fingerprint

    Dive into the research topics of 'Reassembling and testing of a high-precision heat capacity drop calorimeter. Heat capacity of some polyphenyls at T = 298.15 K'. Together they form a unique fingerprint.

    Cite this