MANTIS: A real-time quantitative multispectral imaging system for fusion plasmas

A. Perek (Corresponding author), W.A.J. Vijvers, Y. Andrebe, I.G.J. Classen, B.P. Duval, C. Galperti, J.R. Harrison, B.L. Linehan, T. Ravensbergen, K. Verhaegh, M.R. de Baar

Research output: Contribution to journalArticleAcademicpeer-review

49 Citations (Scopus)
252 Downloads (Pure)

Abstract

This work presents a novel, real-time capable, 10-channel Multispectral Advanced Narrowband Tokamak Imaging System installed on the TCV tokamak, MANTIS. Software and hardware requirements are presented together with the complete system architecture. The image quality of the system is assessed with emphasis on effects resulting from the narrowband interference filters. Some filters are found to create internal reflection images that are correlated with the filters' reflection coefficient. This was measured for selected filters where significant absorption (up to 65% within ∼70 nm of the filter center) was measured. The majority of this was attributed to the filter's design, and several filters' performance is compared. Tailored real-time algorithms exploiting the system's capabilities are presented together with benchmarks comparing polling and event based synchronization. The real-time performance is demonstrated with a density ramp discharge performed on TCV. The behavior of spectral lines' emission from different plasma species and their interpretation are qualitatively described.

Original languageEnglish
Article number123514
Number of pages12
JournalReview of Scientific Instruments
Volume90
Issue number12
DOIs
Publication statusPublished - 1 Dec 2019

Fingerprint

Dive into the research topics of 'MANTIS: A real-time quantitative multispectral imaging system for fusion plasmas'. Together they form a unique fingerprint.

Cite this