Background and purpose: Hyperthermia, the elevation of target temperature to 39–44 °C, is monitored using temperature probes. However, these provide limited spatial information, sampling only a few discrete locations. Magnetic resonance (MR) thermometry currently offers an option for three-dimensional (3D) temperature monitoring during hyperthermia. This study compares and correlates temperatures measured by intraluminal probes with MR-based temperatures in (1) the anatomical region containing the intraluminal probes and (2) the hyperthermia target volume (HTV), located at a distance from the probes and representing the primary region of clinical interest. Methods: Thirteen locally advanced cervical cancer (LACC) patients treated with radiotherapy and hyperthermia were included. Hyperthermia was monitored using intraluminal probes and MR thermometry. MR-based temperatures were compared to intraluminal probe temperatures. Repeated measures correlation was applied to correlate probe and MR-based temperatures in the HTV across all data and on a patient-specific basis. Results: MR-based temperatures at probe locations showed good agreement with probe measurements (median absolute error ≤ 0.7 °C). In the HTV, MR-based temperatures deviated by a median absolute error of 0.5 °C from probe temperatures. Repeated measures correlations (rrm) between MR and probe-based HTV temperatures ranged from 0.74 to 0.79 across all data and 0.64–0.96 on a patient-specific basis. Conclusions: MR thermometry demonstrated promising performance for retrospective evaluation of temperature distributions in the HTV. While its current reliability for real-time treatment guidance remains limited, our results support further development towards broader clinical implementation in hyperthermia.

Original languageEnglish
Article number100812
Number of pages8
JournalPhysics and Imaging in Radiation Oncology
Volume35
DOIs
Publication statusPublished - Jul 2025

Bibliographical note

Publisher Copyright:
© 2025 The Author(s)

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Hyperthermia
  • Locally advanced cervical cancer
  • Magnetic resonance thermometry
  • Magnetic resonance-guided hyperthermia
  • Probe thermometry
  • Repeated measures correlation
  • Target volume temperature

Fingerprint

Dive into the research topics of 'Magnetic resonance thermometry in the target volume versus intraluminal probe thermometry for hyperthermia treatment monitoring'. Together they form a unique fingerprint.

Cite this