Abstract
One of the major challenges in the design of multichannel arrays for UHF MRI is minimizing coupling between elements. Shielded-coaxial-cable (SCC)
coils have been recently proposed as highly decoupled elements per se: however, the decoupling mechanism of SCC coils is still not determined. In this work, we
examined the relationship between coupling, surface current density uniformity, and the total current on coil conductor through measuring and simulating three
different 2-coils arrays. While no correlation was observed between surface current density uniformity and interelement coupling, a negative correlation was
observed between the coupling and the total current on coil conductor.
coils have been recently proposed as highly decoupled elements per se: however, the decoupling mechanism of SCC coils is still not determined. In this work, we
examined the relationship between coupling, surface current density uniformity, and the total current on coil conductor through measuring and simulating three
different 2-coils arrays. While no correlation was observed between surface current density uniformity and interelement coupling, a negative correlation was
observed between the coupling and the total current on coil conductor.
Original language | English |
---|---|
Pages | P-082 |
Number of pages | 1 |
Publication status | Published - 2022 |
Event | 14th ISMRM Benelux Chapter Meeting 2022 - MECC, Maastricht, Netherlands Duration: 25 Apr 2022 → 25 Apr 2022 Conference number: 14 |
Conference
Conference | 14th ISMRM Benelux Chapter Meeting 2022 |
---|---|
Abbreviated title | ISMRM |
Country/Territory | Netherlands |
City | Maastricht |
Period | 25/04/22 → 25/04/22 |