Abstract
We propose a shielded loop coil design for operation as a receive array element at 7T. The proposed coil geometry provides high decoupling between adjacent and non-adjacent loop elements without overlapping or preamplifier decoupling. A four-coil receive array was constructed for high resolution imaging of the knee at 7T. The coil coupling between the elements for in-vivo measurements was lower than -26 dB. Images with an isotropic resolution of 0.7 x 0.7 x 0.7 mm were acquired in five minutes.
| Original language | English |
|---|---|
| Title of host publication | 11th Annual Meeting ISMRM Benelux Chapter |
| Publication status | Published - 2019 |
| Externally published | Yes |
| Event | 11th ISMRM Benelux Chapter Meeting 2019 - Stadsgehoorland, Leiden, Netherlands Duration: 17 Jan 2019 → 17 Jan 2019 http://www.ismrm-benelux.org/annual-meetings/meeting-2019/ |
Conference
| Conference | 11th ISMRM Benelux Chapter Meeting 2019 |
|---|---|
| Country/Territory | Netherlands |
| City | Leiden |
| Period | 17/01/19 → 17/01/19 |
| Internet address |