Skip to main navigation Skip to search Skip to main content

Evaluation of a new double-lumen endobronchial tube with an integrated camera (VivaSight-DL): A prospective multicentre observational study

  • E.M. Koopman
  • , M. Barak
  • , E. Weber
  • , M.J.A. Valk
  • , R.T.I. de Schepper
  • , R.A. Bouwman
  • , J.M. Huitink

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

The VivaSight-DL is a new single-use double-lumen endobronchial tube with an integrated camera. We studied this device in 151 consecutive patients scheduled for elective thoracic surgery in four different hospitals. Endobronchial intubation was successful in 148 patients (98%) (95% CI 94-99%). Median (IQR [range]) endobronchial intubation time was 59 (47-82 [17-932]) s and lung isolation was successfully achieved in 147 (99%) patients (95% CI 96-99%). A fibreoptic bronchoscope was required to assist endobronchial tube placement in 19 (13%) patients (95% CI 8-19%). Sore throat was reported by 37 (25%) patients (95% CI 18-33%), but no major complications were observed. We have reported the successful use of the VivaSight double-lumen tube for endobronchial intubation in a multicentre observational trial.

Original languageEnglish
Pages (from-to)962-968
Number of pages7
JournalAnaesthesia
Volume70
Issue number8
DOIs
Publication statusPublished - Aug 2015
Externally publishedYes

Fingerprint

Dive into the research topics of 'Evaluation of a new double-lumen endobronchial tube with an integrated camera (VivaSight-DL): A prospective multicentre observational study'. Together they form a unique fingerprint.

Cite this