Abstract
The VivaSight-DL™ is a new single-use double-lumen endobronchial tube with an integrated camera. We studied this device in 151 consecutive patients scheduled for elective thoracic surgery in four different hospitals. Endobronchial intubation was successful in 148 patients (98%) (95% CI 94-99%). Median (IQR [range]) endobronchial intubation time was 59 (47-82 [17-932]) s and lung isolation was successfully achieved in 147 (99%) patients (95% CI 96-99%). A fibreoptic bronchoscope was required to assist endobronchial tube placement in 19 (13%) patients (95% CI 8-19%). Sore throat was reported by 37 (25%) patients (95% CI 18-33%), but no major complications were observed. We have reported the successful use of the VivaSight double-lumen tube for endobronchial intubation in a multicentre observational trial.
| Original language | English |
|---|---|
| Pages (from-to) | 962-968 |
| Number of pages | 7 |
| Journal | Anaesthesia |
| Volume | 70 |
| Issue number | 8 |
| DOIs | |
| Publication status | Published - Aug 2015 |
| Externally published | Yes |
Fingerprint
Dive into the research topics of 'Evaluation of a new double-lumen endobronchial tube with an integrated camera (VivaSight-DL™): A prospective multicentre observational study'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver