Abstract
Objectives: The ELANA® Heart Bypass creates a standardized sutureless anastomosis. Hereby, we investigate the influence of arteriotomy and graft size on coronary hemodynamics. Methods: A computational fluid dynamics (CFD) model was developed. Arteriotomy size (standard 1.43 mm2; varied 0.94 – 3.6 mm2) and graft diameter (standard 2.5 mm; varied 1.5 – 5.0 mm) were independent parameters. Outcome parameters were coronary pressure and flow, and fractional flow reserve (FFR). Results: The current size ELANA (arteriotomy 1.43 mm2) presented an estimated FFR 0.65 (39 mL/min). Enlarging arteriotomy increased FFR, coronary pressure, and flow. All reached a maximum once the arteriotomy (2.80 mm2) surpassed the coronary cross-sectional area (2.69 mm2, i.e. 1.85 mm diameter), presenting an estimated FFR 0.75 (46 mL/min). Increasing graft diameter was positively related to FFR, coronary pressure, and flow. Conclusion: The ratio between the required minimal coronary diameter for application and the ELANA arteriotomy size effectuates a pressure drop that could be clinically relevant. Additional research and eventual lengthening of the anastomosis is advised. Graphical abstract: [Figure not available: see fulltext.].
| Original language | English |
|---|---|
| Pages (from-to) | 916-926 |
| Number of pages | 11 |
| Journal | Journal of Cardiovascular Translational Research |
| Volume | 16 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - Aug 2023 |
Keywords
- Computational fluid dynamics
- Coronary artery bypass grafting
- Fractional flow reserve
- Sutureless coronary anastomoses
Fingerprint
Dive into the research topics of 'Evaluating the Arteriotomy Size of a New Sutureless Coronary Anastomosis Using a Finite Volume Approach'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver