Design of an experimental set-up to study the behavior of a flexible surgical instrument inside an endoscope

J.P. Khatait, D.M. Brouwer, H.M.J.R. Soemers, R.G.K.M. Aarts, J.L. Herder

Research output: Contribution to journalArticleAcademicpeer-review

2 Citations (Scopus)

Abstract

The success of flexible instruments in surgery requires high motion and force fidelity and controllability of the tip. However, the friction and the limited stiffness of such instruments limit the motion and force transmission of the instrument. In a previous study, we developed a flexible multibody model of a surgical instrument inside an endoscope in order to study the effect of the friction, bending and rotational stiffness of the instrument and clearance on the motion hysteresis and the force transmission. In this paper, we present the design and evaluation of an experimental setup for the validation of the flexible multibody model and the characterization of the instruments. A modular design was conceived based on three key functionalities: the actuation from the proximal end, the displacement measurement of the distal end, and the measurement of the interaction force. The exactly constrained actuation module achieves independent translation and rotation of the proximal end. The axial displacement and the rotation of the distal end are measured contactless via a specifically designed air bearing guided cam through laser displacement sensors. The errors in the static measurement are 15 lm in translation and 0.15 deg in rotation. Six 1-DOF load cell modules using flexures measure the interaction forces and moments with an error of 0.8% and 2.5%, respectively. The achieved specifications allow for the measurement of the characteristic behavior of the instrument inside a curved rigid tube and the validation of the flexible multibody model. [DOI: 10.1115/1.4024660].

Original languageEnglish
Article number031004
JournalJournal of Medical Devices
Volume7
Issue number3
DOIs
Publication statusPublished - 3 Jul 2013

Fingerprint

Endoscopy
Endoscopes
Surgical Instruments
Research Design
Friction
Lasers
Air
Bearings (structural)
Stiffness
Displacement measurement
Cams
Controllability
Surgery
Hysteresis
Specifications
Sensors

Keywords

  • Design
  • Evaluation
  • Experimental setup
  • Flexible wire
  • Surgical instrument

Cite this

@article{e3a4cfc2406f4f0490d556cae69a8715,
title = "Design of an experimental set-up to study the behavior of a flexible surgical instrument inside an endoscope",
abstract = "The success of flexible instruments in surgery requires high motion and force fidelity and controllability of the tip. However, the friction and the limited stiffness of such instruments limit the motion and force transmission of the instrument. In a previous study, we developed a flexible multibody model of a surgical instrument inside an endoscope in order to study the effect of the friction, bending and rotational stiffness of the instrument and clearance on the motion hysteresis and the force transmission. In this paper, we present the design and evaluation of an experimental setup for the validation of the flexible multibody model and the characterization of the instruments. A modular design was conceived based on three key functionalities: the actuation from the proximal end, the displacement measurement of the distal end, and the measurement of the interaction force. The exactly constrained actuation module achieves independent translation and rotation of the proximal end. The axial displacement and the rotation of the distal end are measured contactless via a specifically designed air bearing guided cam through laser displacement sensors. The errors in the static measurement are 15 lm in translation and 0.15 deg in rotation. Six 1-DOF load cell modules using flexures measure the interaction forces and moments with an error of 0.8{\%} and 2.5{\%}, respectively. The achieved specifications allow for the measurement of the characteristic behavior of the instrument inside a curved rigid tube and the validation of the flexible multibody model. [DOI: 10.1115/1.4024660].",
keywords = "Design, Evaluation, Experimental setup, Flexible wire, Surgical instrument",
author = "J.P. Khatait and D.M. Brouwer and H.M.J.R. Soemers and R.G.K.M. Aarts and J.L. Herder",
year = "2013",
month = "7",
day = "3",
doi = "10.1115/1.4024660",
language = "English",
volume = "7",
journal = "Journal of Medical Devices",
issn = "1932-6181",
publisher = "American Society of Mechanical Engineers",
number = "3",

}

Design of an experimental set-up to study the behavior of a flexible surgical instrument inside an endoscope. / Khatait, J.P.; Brouwer, D.M.; Soemers, H.M.J.R.; Aarts, R.G.K.M.; Herder, J.L.

In: Journal of Medical Devices, Vol. 7, No. 3, 031004, 03.07.2013.

Research output: Contribution to journalArticleAcademicpeer-review

TY - JOUR

T1 - Design of an experimental set-up to study the behavior of a flexible surgical instrument inside an endoscope

AU - Khatait, J.P.

AU - Brouwer, D.M.

AU - Soemers, H.M.J.R.

AU - Aarts, R.G.K.M.

AU - Herder, J.L.

PY - 2013/7/3

Y1 - 2013/7/3

N2 - The success of flexible instruments in surgery requires high motion and force fidelity and controllability of the tip. However, the friction and the limited stiffness of such instruments limit the motion and force transmission of the instrument. In a previous study, we developed a flexible multibody model of a surgical instrument inside an endoscope in order to study the effect of the friction, bending and rotational stiffness of the instrument and clearance on the motion hysteresis and the force transmission. In this paper, we present the design and evaluation of an experimental setup for the validation of the flexible multibody model and the characterization of the instruments. A modular design was conceived based on three key functionalities: the actuation from the proximal end, the displacement measurement of the distal end, and the measurement of the interaction force. The exactly constrained actuation module achieves independent translation and rotation of the proximal end. The axial displacement and the rotation of the distal end are measured contactless via a specifically designed air bearing guided cam through laser displacement sensors. The errors in the static measurement are 15 lm in translation and 0.15 deg in rotation. Six 1-DOF load cell modules using flexures measure the interaction forces and moments with an error of 0.8% and 2.5%, respectively. The achieved specifications allow for the measurement of the characteristic behavior of the instrument inside a curved rigid tube and the validation of the flexible multibody model. [DOI: 10.1115/1.4024660].

AB - The success of flexible instruments in surgery requires high motion and force fidelity and controllability of the tip. However, the friction and the limited stiffness of such instruments limit the motion and force transmission of the instrument. In a previous study, we developed a flexible multibody model of a surgical instrument inside an endoscope in order to study the effect of the friction, bending and rotational stiffness of the instrument and clearance on the motion hysteresis and the force transmission. In this paper, we present the design and evaluation of an experimental setup for the validation of the flexible multibody model and the characterization of the instruments. A modular design was conceived based on three key functionalities: the actuation from the proximal end, the displacement measurement of the distal end, and the measurement of the interaction force. The exactly constrained actuation module achieves independent translation and rotation of the proximal end. The axial displacement and the rotation of the distal end are measured contactless via a specifically designed air bearing guided cam through laser displacement sensors. The errors in the static measurement are 15 lm in translation and 0.15 deg in rotation. Six 1-DOF load cell modules using flexures measure the interaction forces and moments with an error of 0.8% and 2.5%, respectively. The achieved specifications allow for the measurement of the characteristic behavior of the instrument inside a curved rigid tube and the validation of the flexible multibody model. [DOI: 10.1115/1.4024660].

KW - Design

KW - Evaluation

KW - Experimental setup

KW - Flexible wire

KW - Surgical instrument

UR - http://www.scopus.com/inward/record.url?scp=84885044164&partnerID=8YFLogxK

U2 - 10.1115/1.4024660

DO - 10.1115/1.4024660

M3 - Article

AN - SCOPUS:84885044164

VL - 7

JO - Journal of Medical Devices

JF - Journal of Medical Devices

SN - 1932-6181

IS - 3

M1 - 031004

ER -