Abstract
Catheter ablation is an increasingly important curative procedure for atrial fibrillation. Knowledge of the local wall thickness is essential to determine the proper ablation energy. This paper presents the first semi-automatic atrial wall thickness measurement method for ablation guidance. It includes both endocardial and epicardial atrial wall segmentation on CT image data. Segmentation is based on active contours, Otsu's multiple threshold method and hysteresis thresholding. Segmentation results were compared to contours manually drawn by two experts, using repeated measures analysis of variance. The root mean square differences between the semi-automatic and the manually drawn contours were comparable to intra-observer variation (endocardium: p = 0.23, epicardium: p = 0.18). Mean wall thickness difference is significant between one of the experts on one side, and the presented method and the other expert on the other side (p