Abstract
Objective. To evaluate the
feasibility of closed-loop blood
pressure control during cardiac
surgery.
Design: A closed-loop system regulated
peroperative hypertension by
controlling the infusion rate of the
vasodilator nitroglycerin (NTG).
The controller consisted of a regulator
which was monitored by
a supervisory computer program.
Mean arterial pressure (MAP) was
calculated every 5 s from measurements
of the radial artery pressure
signal. The regulator calculated an
NTG infusion rate with each new
MAP measurement. The supervisory
computer program monitored
the regulator's actions and
adapted or overruled the regulator
when required.
Setting. The cardiac surgery operating
room.
Patients." 46 patients who were
scheduled for cardiac surgery and
who developed peroperative hypertension.
Interventions. Patients were scheduled
for either bypass or valve
replacement surgery. The closedloop
system was used to control
hypertension before and after cardiopulmonary
bypass. The use of
the closed-loop system did not
require deviation from the protocol
normally used during cardiac surgery.
All patients received standard
continuous anaesthesia with
opioids.
Measurements and results: Initial
automatic control was achieved in
9.4 (4.1 SD) rain. The percentage of
time that MAP remained in a range
around the target MAP of + 10
and _ 20 mmHg was 74 and 94%,
respectively. The mean NTG infusion
rate while MAP was within
5 mmHg of target MAP was 1.14
(0.84 SD) ggkg -1 rain -1. Target
MAP was set between 65 and
90 mmHg. There was a small group
of patients (6 out of 46) who did not
respond to NTG and required
alternative drug therapy.
Conclusions: The controller provided
fast and stable control in all
patients. The expert knowledge
implemented through the supervisory
computer program enabled the
controller to respond adequately to
the rapid changes in arterial pressures
commonly associated with
cardiac surgery. We conclude that
closed-loop control of arterial pressure
is feasible not only in the
cardiac surgical care unit but also
during cardiac surgery.
| Original language | English |
|---|---|
| Pages (from-to) | 688-693 |
| Number of pages | 6 |
| Journal | Intensive Care Medicine |
| Volume | 22 |
| DOIs | |
| Publication status | Published - 1996 |
Fingerprint
Dive into the research topics of 'Automated infusion of nitroglycerin to control arterial hypertension during cardiac surgery'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver