Adapted ST analysis during labor: relative versus absolute ST events, a case-control study

Alexandra D.J. Hulsenboom (Corresponding author), Daisy A.A. van der Woude, Martina M. Porath, Anneke Kwee, Rik Vullings, S. Guid Oei, Judith O.E.H. van Laar

Research output: Contribution to journalArticleAcademicpeer-review

37 Downloads (Pure)

Abstract

BACKGROUND: The value of ST analysis of the fetal electrocardiogram during labor to lower asphyxia and cesarean section rates is uncertain. Physiological variation of the electrical heart axis between fetuses may explain false alarms in conventional ST analysis (absolute ST analysis). ST events (alarms) based on relative T/QRS rises (relative ST analysis) correct for this variation and may improve diagnostic accuracy of ST analysis.

AIMS: To compare the diagnostic accuracy of absolute and relative ST analysis with regard to fetal acidemia.

STUDY DESIGN: Retrospective case-control study.

SUBJECTS: 20 healthy women with an uncomplicated pregnancy monitored with ST analysis during labor: 10 cases (umbilical cord artery pH < 7.05) and 10 controls (pH > 7.20).

OUTCOME MEASURES: Sensitivity, specificity, positive and negative likelihood ratio.

RESULTS: In 16 of the 20 patients a total of 54 absolute ST events were reported. Two reviewers classified the cardiotocograms; in cases 29% of the absolute ST events were significant, in the controls it was 19%. Relative ST analysis versus absolute ST analysis showed a sensitivity of 90% (55-100%) vs. 70% (35-93%), a specificity of 100% (69-100%) vs. 70% (35-93%), a positive likelihood ratio of infinity vs. 2.3 (0.8-6.5), a negative likelihood ratio of 0.1 (0.0-0.6) vs. 0.4 (0.2-1.2), and diagnostic odds ratio of infinity vs. 5.4 (0.8-36.9). McNemar showed no statistical significant difference between the sensitivity and specificity of the methods.

CONCLUSIONS: We observed higher positive and lower negative likelihood ratios for relative ST analysis in comparison to absolute ST analysis. In this small study we found no statistical difference. Relative ST analysis should be studied in a larger study.

Original languageEnglish
Pages (from-to)7375-7380
Number of pages6
JournalJournal of Maternal-Fetal and Neonatal Medicine
Volume35
Issue number25
DOIs
Publication statusPublished - Dec 2022

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Keywords

  • Diagnostic accuracy
  • fetal electrocardiography
  • fetal monitoring
  • metabolic acidosis
  • ST event
  • T/QRS
  • Heart Rate, Fetal
  • Fetal Monitoring/methods
  • Humans
  • Case-Control Studies
  • Pregnancy
  • Cesarean Section
  • Labor, Obstetric
  • Electrocardiography/methods
  • Cardiotocography/methods
  • Female
  • Retrospective Studies