Accelerated high-frame-rate mouse heart cine-MRI using compressed sensing reconstruction

A.G.A. Mohamed, B.F. Coolen, D. Abdurrachim, R.M. Pires da Silva Castro, L.M.J. Florack, J.J. Prompers, K. Nicolaij, G.J. Strijkers

Research output: Contribution to journalArticleAcademicpeer-review

20 Citations (Scopus)

Abstract

We introduce a new protocol to obtain very high-frame-rate cinematographic (Cine) MRI movies of the beating mouse heart within a reasonable measurement time. The method is based on a self-gated accelerated fast low-angle shot (FLASH) acquisition and compressed sensi ng reconstruction. Key to our approach is that we exploit the stochastic nature of the retrospective triggering acquisition scheme to produce an undersampled and random k–t space filling that allows for compressed sensing reconstruction and acceleration. As a standard, a self-gated FLASH sequence with a total acquisition time of 10 min was used to produce single-slice Cine movies of seven mouse hearts with 90 frames per cardiac cycle. Two times (2×) and three times (3×) k–t space undersampled Cine movies were produced from 2.5- and 1.5-min data acquisitions, respectively. The accelerated 90-frame Cine movies of mouse hearts were successfully reconstructed with a compressed sensing algorithm. The movies had high image quality and the undersampling artifacts were effectively removed. Left ventricular functional parameters, i.e. end-systolic and end-diastolic lumen surface areas and early-to-late filling rate ratio as a parameter to evaluate diastolic function, derived from the standard and accelerated Cine movies, were nearly identical.
Original languageEnglish
Pages (from-to)451-457
JournalNMR in Biomedicine
Volume26
Issue number4
DOIs
Publication statusPublished - 2013

Fingerprint

Dive into the research topics of 'Accelerated high-frame-rate mouse heart cine-MRI using compressed sensing reconstruction'. Together they form a unique fingerprint.

Cite this