A novel CGC-CGC-MS configuration for the analysis of dioxins and furans

F. David, P.J.F. Sandra, A. Hoffman, J. Gerstel

Research output: Chapter in Book/Report/Conference proceedingConference contributionAcademicpeer-review

43 Citations (Scopus)

Abstract

The analysis of dioxins and furans on a novel multidimensional (CGC - CGC - MS) configuration is presented. The system is based on flow controlled column switching, allowing the simultaneous isomeric separation of tetrachloro-, pentachloro- and hexachloro-substituted dibenzodioxins and dibenzofurans on a polar column and the analysis of heptachloro- and octachloro-substituted dibenzodioxins and dibenzofurans on an apolar column. PI'V injection in the solvent vent mode allows analysis of large sample sizes approaching in this way the sensitivity of CGC - high resolution MS.
Original languageEnglish
Title of host publicationProceedings of the 14th International Symposium on Capillary Chromatography, 25-29 May,1992, Baltimore, USA
EditorsP. Sandra, M.L. Lee
Place of PublicationHeidelberg
PublisherHuethig
Pages671-679
Publication statusPublished - 1992
Event14th International Symposium on Capillary Chromatography - Stouffer Harborplace Hotel, Baltimore, United States
Duration: 25 May 199229 May 1992

Conference

Conference14th International Symposium on Capillary Chromatography
CountryUnited States
CityBaltimore
Period25/05/9229/05/92