A low power configurable bio-impedance spectroscopy (BIS) ASIC with simultaneous ECG and respiration recording functionality

J. Xu, P.J.A. Harpe, J. Pettine, C. Van Hoof, R.F. Yazicioglu

Research output: Chapter in Book/Report/Conference proceedingConference contributionAcademicpeer-review

28 Citations (Scopus)
3 Downloads (Pure)

Abstract

This paper presents a power-efficient ASIC for bio-impedance spectroscopy, as well as ECG and respiration recording. The ASIC includes a wideband stimulation current source with a pseudo-random binary sequence, a low-noise instrumentation amplifier, and a low power 12b SAR ADC. This ASIC is able to measure bio-impedances from 1ω to 10kω with 0.1ω resolution, and covers a frequency range up to 125kHz. Furthermore, the ASIC can simultaneously record ECG and respiration while consuming only 31μW from a 1.8V supply.

Original languageEnglish
Title of host publicationESSCIRC 2015 - Proceedings of the 41st European Solid-State Circuits Conference, 14-18 September 2015, Graz, Austria
PublisherIEEE Computer Society
Pages396-399
Number of pages4
VolumeBrussels
ISBN (Electronic)9781467374705
DOIs
Publication statusPublished - 30 Oct 2015
Event41st European Solid-State Circuits Conference (ESSCIRC 2015) - Graz, Austria
Duration: 14 Sept 201518 Sept 2015
Conference number: 41

Conference

Conference41st European Solid-State Circuits Conference (ESSCIRC 2015)
Country/TerritoryAustria
CityGraz
Period14/09/1518/09/15

Fingerprint

Dive into the research topics of 'A low power configurable bio-impedance spectroscopy (BIS) ASIC with simultaneous ECG and respiration recording functionality'. Together they form a unique fingerprint.

Cite this