A BGO Detector for Positron Emission Profiling in catalysis

A.V.G. Mangnus, L.J. IJzendoorn, van, J.J.M. de Goeij, R.H. Cunningham, R.A. Santen, van, M.J.A. Voigt, de

Research output: Contribution to journalArticleAcademicpeer-review

22 Citations (Scopus)
57 Downloads (Pure)

Abstract

As part of a project to study the reaction kinetics in catalysts, a detector system has been designed and built. The detector will measure in one dimension the activity distribution of positron emitters in catalyst reactors under operational conditions as a function of time. The detector consists of two arrays of ten EGO crystals each and has the flexibility to measure with high sensitivity the activity profile in various reactor sizes; the position resolution that can be reached is 3 mm
Original languageEnglish
Pages (from-to)649-652
JournalNuclear Instruments and Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms
Volume99
Issue number1-4
DOIs
Publication statusPublished - 1995

Fingerprint Dive into the research topics of 'A BGO Detector for Positron Emission Profiling in catalysis'. Together they form a unique fingerprint.

Cite this